Gluclas 2.0 - A mobile app for clinical decision support during glucose clamps.
Giovanni Tubaro, Bettina Weber, Giacomo Cappon, David Studer, Michele Schiavon, David Herzig, Chiara Dalla Man, Lia Bally,
Department of Information Engineering, University of Padova, Padova, Italy. · Department of Diabetes, Endocrinology, Nutritional Medicine and Metabolism, Inselspital, Bern University Hospital and University of Bern, Bern, Switzerland; Graduate School for Cellular and Biomedical Sciences, University of Bern, 3012, Bern, Switzerland.
Key points
From the paper's conclusion- Gluclas 2.0 achieved satisfactory reference-tracking performance in simulations conducted in silico and showed promising preliminary feasibility in two human ISO-IV clamps. While broader clinical validation is required, these findings support its potential as a low-cost decision-support tool for standardizing ISO-IV clamp procedures and reducing operator-dependent variability.
Abstract
BACKGROUND AND OBJECTIVE: Glucose clamps are valuable yet complex experimental techniques used in metabolic research. Among these, ISOglycemic IntraVenous (ISO-IV) clamps are used to assess the incretin effect, that is, the enhanced insulin response following nutrient intake compared to an equivalent glucose increase induced intravenously. ISO-IV clamps require replicating the glucose profile of an oral glucose tolerance test through IV infusion, making them particularly challenging and prone to operator variability. This study presents Gluclas 2.0, a mobile application designed to assist clinicians in modulating glucose infusion rates (GIR) during ISO-IV clamps. METHOD: Gluclas 2.0 integrates a Proportional-Integral-Derivative controller with a personalized feedforward component for proactive GIR control. The system is implemented in a mobile app featuring a user-friendly interface. It was tuned and validated in silico using 87 virtual prediabetic subjects, simulating realistic conditions including measurement noise and sampling delays. Preliminary in vivo testing was also conducted on two human subjects. RESULTS: In silico validation yielded a population-wide Mean Absolute Percentage Error (MAPE) of 4.12%, below the 5% threshold for effective clamping. Preliminary in vivo tests produced results within the range observed in silico. CONCLUSION: Gluclas 2.0 achieved satisfactory reference-tracking performance in simulations conducted in silico and showed promising preliminary feasibility in two human ISO-IV clamps. While broader clinical validation is required, these findings support its potential as a low-cost decision-support tool for standardizing ISO-IV clamp procedures and reducing operator-dependent variability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Closed-loop insulin delivery for glycaemic control in hospitalised and perioperative adults: A systematic review and meta-analysis of randomised controlled trials.Diabetes research and clinical practice · 2026
- Development and external validation of machine learning models for predicting glucose trajectories after continuous intravenous insulin therapy in hyperglycemic crises.Journal of diabetes investigation · 2026
- Controlled In Vitro Characterization of the Dynamic Response of Continuous Glucose Monitoring Systems: Adaptation of a Programmable Flow Platform and Decomposition of Dynamic ErrorbioRxiv · 2026
- Closed-Loop Insulin Delivery in Inpatient Care: A Systematic Review and Meta-Analysis of Clinical and Implementation Outcomes.Diabetes, obesity & metabolism · 2026
- A Human Glucoregulatory Model for the Rational Design and Evaluation of Diabetes Therapeutics.ACS pharmacology & translational science · 2026
- Fractional-Order PI Control for Automated Insulin Dosing: A Simulation-Based Comparison with Integer-Order PI Control Across Patient Age GroupsFractal and Fractional · 2026