An advanced hybrid automated insulin delivery system for insulin-dependent diabetes based on successive linearization model predictive control: The UniBE system.
Vihangkumar V Naik, Eleonora Manzoni, Clara Escorihuela-Altaba, Jose Garcia-Tirado
Department of Diabetes, Endocrinology, Nutritional Medicine, and Metabolism, Inselspital, Bern University Hospital and University of Bern, Switzerland; Diabetes Center Berne, Bern, Switzerland. Electronic address: vihangkumar.naik@unibe.ch. · Department of Diabetes, Endocrinology, Nutritional Medicine, and Metabolism, Inselspital, Bern University Hospital and University of Bern, Switzerland; Diabetes Center Berne, Bern, Switzerland. Electronic address: eleonora.manzoni@unibe.ch.
Key points
From the paper's conclusion- The UniBE hAID system maintained stable glycemic regulation across diverse error conditions, highlighting its robustness and adaptability. These findings establish proof-of-concept feasibility and provide a foundation for further refinement, although these do not constitute a full pre-clinical validation. Future work must extend beyond in-silico testing to include more preclinical and clinical studies to rigorously assess safety, efficacy, and real-world performance.
Abstract
BACKGROUND AND OBJECTIVE: Hybrid automated insulin delivery (hAID) systems represent the most advanced therapy for type 1 diabetes (T1D). Current systems rely on linear or linearized models of glucose homeostasis, which may compromise prediction accuracy and, in turn, timely decision-making by the controller. Physiological variability further complicates insulin requirements, underscoring the need for controllers that adapt dynamically and reduce user burden. METHODS: We introduce the University of Bern (UniBE) hAID system, a framework based on successive linearization model predictive control (MPC). The controller integrates basal insulin infusion with the insulin bolus delivery module for meal-related and corrective bolus dosing, adapting bounds in real time to glucose dynamics while accounting for both automated and user-initiated inputs. In-silico evaluation was conducted using the commercial version of the FDA-accepted UVa/Padova metabolic simulator across nine scenarios involving persistent and time-varying errors in meal timing, carbohydrate estimation, and basal insulin profiles. RESULTS: In the baseline scenario, UniBE achieved a median [interquartile ranges] time in range of 97.3 [90.8 99.1]%, with time below range at 0 [0 0.3]% and time above range at 2.4 [0.9 9.2]%. Across perturbation scenarios, median time in range remained between 81.7%-95.0%, with low hypoglycemia incidence, demonstrating resilience to clinically relevant disturbances. CONCLUSION: The UniBE hAID system maintained stable glycemic regulation across diverse error conditions, highlighting its robustness and adaptability. These findings establish proof-of-concept feasibility and provide a foundation for further refinement, although these do not constitute a full pre-clinical validation. Future work must extend beyond in-silico testing to include more preclinical and clinical studies to rigorously assess safety, efficacy, and real-world performance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Design and Simulation of an AI-Powered Artificial Pancreas System With Personalized Glucose Sensing and Insulin Delivery.Artificial organs · 2026
- Hybrid Stochastic–Machine Learning Framework for Postprandial Glucose Prediction in Type 1 DiabetesAlgorithms · 2025
- Tail-Risk-Aware Model Predictive Control for Hypoglycemia Mitigation in Automated Insulin Delivery.Metabolites · 2026
- Glycemic and safety outcomes of the insulin-only bionic pancreas in older adults and individuals with impaired awareness of Hypoglycemia: a post hoc analysis of a randomized pivotal trial.Diabetes research and clinical practice · 2026
- Capturing human-machine interplay in automated insulin delivery systems through day-to-day variability of the automated correction index (ACI-V%).Diabetes research and clinical practice · 2026
- Precision Insulin-Pump Calibration and Optimal Insulin Delivery to Improve Blood Glucose Levels.medRxiv : the preprint server for health sciences · 2026