Key points
From the paper's conclusion- The proposed architecture demonstrates a safety-audited template for combining machine-learning basal prediction with clinically grounded dosing rules. Validation on real outcome data and physical hardware is identified as the next steps toward clinical translation.
Abstract
BACKGROUND: Achieving tight glycemic control in Type 1 diabetes is challenged by non-linear metabolic dynamics and insulin-stacking risk. Artificial intelligence models offer improved personalization, but their opacity raises safety concerns, particularly for basal insulin dosing, which is often treated as static in existing decision-support systems. METHODS: We propose a hybrid closed-loop system combining a data-driven LightGBM regressor for basal dose prediction, trained on the T1D-UOM dataset, with a physiologically grounded, rule-based safety layer at the edge. Basal insulin is predicted from cyclical time encoding and glucose rate-of-change; bolus dosing follows carbohydrate intake and correction requirements. A time-bounded insulin-on-board model, an uncertainty-adaptive post-bolus damping factor, and a predictive low-glucose-suspend rule reproducing the Tandem Basal-IQ protocol together suspend or reduce delivery when residual insulin is high or hypoglycemia is projected. The system was evaluated in silico over 60 continuous days under shadow-mode replay and closed-loop reduced-order simulation, including an ablation with bolus delivery suppressed. RESULTS: Under chronological, within-subject validation, the offline basal estimator achieved R 2 = 0.68 and MAE = 0.21 U; under strict leave-one-subject-out validation on the same real cohort, performance did not generalize to unseen subjects ( R 2 = - 0.49 , MAE = 0.49 U). The hybrid controller maintained Time-in-Range at 62.8% within the reduced-order closed-loop simulation, with no violations of the insulin-on-board suspension or damping rules; suppressing bolus delivery in an ablation within this simulation dropped Time-in-Range to 26.1%, indicating that, within the reduced-order model, bolus delivery is a determinant of simulated glycemic outcome. CONCLUSIONS: The proposed architecture demonstrates a safety-audited template for combining machine-learning basal prediction with clinically grounded dosing rules. Validation on real outcome data and physical hardware is identified as the next steps toward clinical translation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Tail-Risk-Aware Model Predictive Control for Hypoglycemia Mitigation in Automated Insulin Delivery.Metabolites · 2026
- Hybrid Stochastic–Machine Learning Framework for Postprandial Glucose Prediction in Type 1 DiabetesAlgorithms · 2025
- An advanced hybrid automated insulin delivery system for insulin-dependent diabetes based on successive linearization model predictive control: The UniBE system.Computer methods and programs in biomedicine · 2026
- Hybrid Digital Twin Framework for Personalized Diabetes Management Using Mathematical Modelling and Machine Learning.Diagnostics (Basel, Switzerland) · 2026
- Generalized multi task learning framework for glucose forecasting and hypoglycemia detection using simulation to realitynpj Digital Medicine · 2025
- Safe Data-Driven Insulin Dosage Decision Based on Physiology-Constrained Gaussian Process LearningarXiv · 2026