Method optimization and verification of homovanillic acid (HVA), vanillylmandelic acid (VMA) and 5-hydroxyindoleacetic acid (5-HIAA) on liquid chromatography tandem mass spectrometry (LC-MS/MS).
Eley Badr, John Beattie, Joseph Macri, Josko Ivica
Hamilton Regional Laboratory Medicine Program (HRLMP), Hamilton Health Sciences and St. Joseph's Healthcare Hamilton, 237 Barton Street E, Hamilton, ON L8L 2X2, Canada. · Hamilton Regional Laboratory Medicine Program (HRLMP), Hamilton Health Sciences and St. Joseph's Healthcare Hamilton, 237 Barton Street E, Hamilton, ON L8L 2X2, Canada; Department of Pathology and Molecular Medicine, Faculty of Health Sciences, McMaster University, 1280 Main Street W, Hamilton, ON L8S 4L8, Canada.
Key points
From the paper's conclusion- We successfully optimized and verified the combination of a Chromsystems kit and laboratory developed method for measuring urinary HVA, VMA and 5-HIAA on our Waters LC-MS/MS.
Abstract
INTRODUCTION: Homovanillic acid (HVA) and vanillylmandelic acid (VMA) are used in diagnosis of neuroblastoma. 5-hydroxyindolacetic acid (5-HIAA) is used in diagnosis of intestinal carcinoid syndrome. The aim of this study is to optimize and verify the Chromsystems kit for these cancer biomarkers on our Waters Xevo TQ-S Cronos LC-MS/MS. MATERIALS AND METHODS: The method was optimized and verified for our LC-MS/MS. The verification included simple (SP) and complex precision (CP), lower limit of detection (LLD) and quantification (LLQ), linearity, carryover, bias through measuring external quality assessment (EQA) samples, and patient comparison with another laboratory that already validated the LC-MS/MS method. RESULTS: For two levels of HVA quality control: SP was 1.3% and 1.8%, CP 4.9% and 3.0%; VMA: SP was 1.5% and 1.3%, CP was 3.5% and 1.6%; and 5-HIAA: SP was 0.7% and 0.9%, CP was 5.4% and 3.8%. The LLD and LLQ were 0.587 μmol/L and 1.20 μmol/L, 0.138 μmol/L and 0.65 μmol/L, and 0.0464 μmol/L and 5.03 μmol/L for HVA, VMA and 5-HIAA, respectively. The linearity was 1.20-290.4 μmol/L (HVA), 0.64-910.7 μmol/L (VMA) and 5.03-1103.4 μmol/L (5-HIAA), without carryover for three analytes. The LC-MS/MS method had no significant bias when compared to the EQA assigned values. The patient comparison had Passing-Bablok regression equations for HVA: y = -0.4801 + 1.153×; VMA: y = -0.3784 + 1.139×; and 5-HIAA: y = 0.0749 + 1.140×. The new reference intervals were adjusted for our method using the method comparison. CONCLUSIONS: We successfully optimized and verified the combination of a Chromsystems kit and laboratory developed method for measuring urinary HVA, VMA and 5-HIAA on our Waters LC-MS/MS.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biogenic amines as neuroendocrine tumor markers: limitations and future perspectives.Clinical chemistry and laboratory medicine · 2026
- A 96-well dual-sample-preparation LC-MS/MS method for integrated quantification of eight urinary catecholamines and metabolites.Clinica chimica acta; international journal of clinical chemistry · 2026
- Targeted LC-MS/MS Analysis of Malonic Acid Biomarkers in Clinical Laboratory Applications.Advances in experimental medicine and biology · 2026
- Eco‐Friendly LC–MS/MS Determination of N ‐Nitroso Vildagliptin Amide, a Nitrosamine Carcinogenic Impurity, in Vildagliptin Drug Substance Using Advanced Greenness MetricsSEPARATION SCIENCE PLUS · 2026
- Development and validation of dispersive liquid-liquid microextraction coupled with LC-MS/MS for plasma 5-hydroxyindoleacetic acid quantification with application as a biomarker in major depressive disorder.Scientific reports · 2026
- Simultaneous Quantification of Total Homocysteine and Methylmalonic Acid by LC-MS/MS for Comprehensive Assessment of Vitamin B12 and Folate deficiencies.Biochimie · 2026