Development and Validation of an RP‐HPLC Method for the Rapid and Sensitive Quantification of Everolimus in Pharmaceutical and Medical Device Applications
Ashwinkumar Vyas, Md. Juned K. Ahmed
Key points
From the paper's conclusion- This validated HPLC method provides a rapid, accurate, and reliable analytical tool for the routine quality control of everolimus in pharmaceutical and medical device applications.
Abstract
Herein, a rapid, sensitive, and reversed‐phase high‐performance liquid chromatography (RP‐HPLC) method was developed and validated for the quantification of everolimus. The method utilizes a reverse‐phase Zorbex SB C18 column with a mobile phase consisting of acetonitrile and methanol, that requires a total run time of ≤ 30 min. Everolimus was detected at a wavelength of 276 nm that ensures clear separation from its related impurities, which included everolimus A, everolimus B, everolimus C, sirolimus B, and sirolimus C. Method validation was performed in accordance with the International Council for Harmonisation, ICH Q2(R1), validation of analytical procedures guidelines; which assesses critical parameters such as linearity, accuracy, precision, and robustness. The method demonstrated excellent linearity ( R 2 ≥ 0.999) over a concentration range of 1–400 ppm, with recovery rates between 99%–100%. The limit of detection and limit of quantification were established at 0.01 ppm (10 ppb) and 0.1 ppm (100 ppb), respectively. The method has been optimized with the following parameters: UV wavelength at 276 nm, column temperature maintained at 55°C, run time of 30 min, ternary solvent system of methanol–water–acetonitrile maintained at a solvent ratio of 61.5%:26%:12.5%. This validated HPLC method provides a rapid, accurate, and reliable analytical tool for the routine quality control of everolimus in pharmaceutical and medical device applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Response Surface Methodology-driven RP-HPLC-PDA Approach to Analytical Method Development, Validation, and Quantitation of Ixazomib in its Pharmaceutical Formulations.Drug metabolism and bioanalysis · 2026
- Green and lean: A validated RP-HPLC-VWD method for simultaneous determination of hydroquinone, methylparaben, fluocinolone acetonide, and propylparaben.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- Development and Validation of RP-HPLC Method for Simultaneous Assay and Dissolution Quantitative Analysis of Pitavastatin-Fenofibrate Complex Dual-Layered TabletsApplied Sciences · 2026
- Analytical quality-by-design-guided green RP-HPLC method to quantify cyclosporin A and luteolin simultaneously in a combination of two drug solutions and two drug-loaded next-generation emulsions or nanocapsules.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- Central Composite Design Approach Stability-Indicating RP-HPLC Analysis for Simultaneous Estimation of Monoclonal Antibodies in Bulk and Combined Dosage Form.Assay and drug development technologies · 2026
- QbD-driven RP-HPLC method for the simultaneous analysis of dihydropyridines calcium channel blockers in pharmaceuticalsBMC Chemistry · 2025