Diagnostic accuracy of photon-counting detector CT for detection of coronary in-stent restenosis.
Karl Aramouni, Chloe Kharsa, Patrick Sarkis, Alpesh R Shah, Waleed Ahmed
Houston Methodist DeBakey Heart & Vascular Center, 6565 Fannin St Suite 1901, Houston, TX, 77030, USA. · Houston Methodist DeBakey Heart & Vascular Center, 6565 Fannin St Suite 1901, Houston, TX, 77030, USA. Electronic address: wahmed@houstonmethodist.org.
Key points
From the paper's conclusion- PCD-CT demonstrated promising diagnostic performance and a high negative predictive value for detecting significant coronary ISR. Its potential utility for excluding ISR, including in smaller stents, warrants validation in larger prospective multicenter studies using standardized acquisition protocols.
Abstract
BACKGROUND: Photon-counting detector computed tomography (PCD-CT) offers improved spatial resolution and reduced blooming artifacts compared with conventional CT, potentially improving visualization of coronary stents. We evaluated the diagnostic performance of PCD-CT for detecting significant coronary in-stent restenosis (ISR) using invasive coronary angiography (ICA) as the reference standard. METHODS: This retrospective, single-center diagnostic-accuracy study included consecutive adults with prior coronary stenting who underwent clinically indicated PCD-CT followed by ICA between February 2025 and January 2026. The primary outcome was the diagnostic performance of PCD-CT for detecting significant ISR, defined as >50% luminal diameter stenosis, on a per-stent basis. Secondary analyses evaluated patient-level performance and exploratory differences according to stent diameter, length, and location. RESULTS: The study included 52 patients (mean age, 70.5 ± 9.7 years; 83% male) with 69 evaluated stents. Significant ISR was present on ICA in 16 stents. PCD-CT correctly identified 12 of 16 stents with ISR and excluded ISR in 44 of 53 stents without ISR, corresponding to a sensitivity of 75.0%, specificity of 83.0%, positive predictive value of 57.1%, negative predictive value of 91.7%, and area under the curve (AUC) of 0.79 (95% CI, 0.67-0.91). Patient-level sensitivity was 86.7%, specificity was 85.7%, and AUC was 0.86 (95% CI, 0.76-0.97). Diagnostic performance was numerically higher in stents ≥3 mm than in stents <3 mm (AUC, 0.82 versus 0.71), although the difference was not statistically significant (P = 0.42). CONCLUSION: PCD-CT demonstrated promising diagnostic performance and a high negative predictive value for detecting significant coronary ISR. Its potential utility for excluding ISR, including in smaller stents, warrants validation in larger prospective multicenter studies using standardized acquisition protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- In vivo coronary stent evaluation using photon-counting computed tomography: diagnostic performance.European radiology · 2026
- Assessing coronary in-stent restenosis using ultrahigh-resolution photon-counting detector CT: results from a two-center studyEuropean Journal of Radiology · 2026
- Diagnostic Performance of Photon‐Counting Coronary CT Angiography in Patients Undergoing TAVI: A Comparison With Invasive AngiographyEuropean Journal of Clinical Investigation · 2026
- Diagnostic performance and clinical impact of photon-counting detector computed tomography in coronary artery disease.Current opinion in cardiology · 2026
- Comparative analysis of photon-counting and energy-integrating detector CT to identify obstructive coronary artery diseaseEuropean Radiology · 2025
- Impact of Photon-Counting CT Acquisition Modes on Coronary Stenosis Assessment and Radiation Dose: A Phantom and Clinical Evaluation.Zeitschrift fur medizinische Physik · 2026