Impact of Photon-Counting CT Acquisition Modes on Coronary Stenosis Assessment and Radiation Dose: A Phantom and Clinical Evaluation.
Thomas Stein, Friederike Lang, Fabian Bamberg, Christopher L Schlett, Jakob Weiss, Philipp Ruile, Jana Taron
Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany. Electronic address: thomas.stein@uniklinik-freiburg.de. · Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
Key points
From the paper's conclusion- While UHR acquisition provides highest image quality in PCD-CT cCTA, lower-dose acquisition modes achieve comparable diagnostic accuracy for coronary stenosis detection. Protocol selection should therefore be individualized, and prospective controlled studies are required.
Abstract
INTRODUCTION: Photon-counting detector computed tomography (PCD-CT) offers high spatial resolution and dose-efficient coronary CT angiography (cCTA), but protocol selection, especially optimal acquisition mode, requires balancing spacial resolution, temporal robustness and radiation exposure. This study evaluated various PCD-CT acquisition modes in terms of image quality, blooming, stenosis grading, diagnostic accuracy, and radiation dose using phantom and clinical cCTA. METHODS: First-generation dual-source PCD-CT was used with four acquisition modes: high-pitch spiral with prospective ECG-gating (Flash), Sequence, Spiral, and retrospective ultra-high-resolution (UHR) spiral mode. Phantom experiments used six 3D-printed coronary artery models with stenoses ranging from 21% to 73%. In a retrospective clinical study, 41 cCTA examinations were analyzed. Qualitative and quantitative image assessments used invasive coronary angiography as reference standard for diagnostic accuracy. Patient characteristics and radiation doses were recorded. RESULTS: UHR demonstrated the highest image quality and the lowest quantitative blooming in phantom and clinical data (clinical image quality score: UHR 4.12 ± 0.99 vs. Flash 3.00 ± 0.86; p = 0.002). Diagnostic accuracy varied across protocols for detecting ≥50% and ≥70% stenoses was high across all modes (e.g., Flash: 76.5% each; Sequence: 100% and 57.1%; UHR: 64.7% and 82.4%). Radiation dose was highest in UHR (26.44 ± 9.76 mSv) compared with Sequence (13.65 ± 8.86 mSv) and Flash (1.40 ± 0.32 mSv); all p ≤ 0.011. CONCLUSION: While UHR acquisition provides highest image quality in PCD-CT cCTA, lower-dose acquisition modes achieve comparable diagnostic accuracy for coronary stenosis detection. Protocol selection should therefore be individualized, and prospective controlled studies are required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic performance and clinical impact of photon-counting detector computed tomography in coronary artery disease.Current opinion in cardiology · 2026
- Ultra-high resolution photon-counting detector coronary CT angiography: diagnostic accuracy in patients with high Agatston scoresEuropean Radiology · 2025
- Spectral ultrahigh-resolution photon-counting CT for coronary stent imaging: evaluation in a dynamic phantomEuropean Radiology Experimental · 2025
- Diagnostic accuracy of photon-counting detector CT for detection of coronary in-stent restenosis.Cardiovascular revascularization medicine : including molecular interventions · 2026
- Photon-counting computed tomography: a revolution in cardiac imagingEuropean Heart Journal · 2025
- In vivo coronary stent evaluation using photon-counting computed tomography: diagnostic performance.European radiology · 2026