Cardiac output monitoring using an automatic arm cuff device in surgical patients.
Mahdi Jazini, Hadi Daher, Ravinder Kumar, Vishaal Dhamotharan, Abhijeet M Sathe, Yaroslava Longhitano, Ezeldeen Abuelkasem, Anand Chandrasekhar,
Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA, USA. · Department of Anesthesiology and Perioperative Medicine, University of Pittsburgh School of Medicine, Pittsburgh, PA, USA.
Key points
From the paper's conclusion- These initial findings suggest potential for convenient CO monitoring using an automatic arm cuff device in surgical patients.
Abstract
BACKGROUND: Pulse contour devices estimate cardiac output (CO) from an invasive arterial blood pressure waveform. However, arterial catheters are generally used only in higher-risk patients. By contrast, automatic arm cuff devices are used for noninvasive BP monitoring in almost all hospital patients. Furthermore, these devices measure cuff pressure oscillations, which are directly related to pulsatile blood volume rather than BP. We hypothesised that automatic arm cuff devices can offer an effective means to monitor CO. METHODS: A custom-built cuff device was used to measure cuff pressure and make a unique measurement of the volume of air pumped into and out of the cuff during linear deflation. Cuff device measurements, an invasive BP waveform from a radial artery catheter, and reference CO from pulmonary artery catheter bolus thermodilution were obtained during interventions from 24 liver transplant and 10 cardiac surgery patients (191 measurement sets). A basic analysis was applied to estimate CO in L min-1 as the product of heart rate, the maximum cuff pressure oscillation amplitude, the cuff-arm system compliance (derived from the cuff pressure-volume relationship), and the body-to-arm surface area ratio (derived from anthropometric information). RESULTS: The arm cuff-based method yielded a correlation of 0.61 (95% confidence interval 0.37-0.79) and a concordance rate of 83% (75%-91%) against the reference CO. For comparison, classic pulse contour analysis of the invasive BP waveform produced a correlation of 0.59 (0.37-0.75) and a concordance rate of 81% (72%-90%). CONCLUSIONS: These initial findings suggest potential for convenient CO monitoring using an automatic arm cuff device in surgical patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of Minimally Invasive and Noninvasive Hemodynamic Monitoring in Adolescents: The Pediatric Cardiac Output Monitoring Observational (POGO) Study.Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies · 2026
- Validation study on cuffless ambulatory blood pressure monitoring method using bio-impedance, impedance cardiography and electrocardiography.European journal of applied physiology · 2026
- Accuracy of cuffless blood pressure devices compared with invasive arterial line monitoring in high-acuity settings: a random-effects meta-analysisEuropean Heart Journal · 2026
- Utilization of hands-free carotid artery Doppler ultrasound for noninvasive central venous pressure measurement: a prospective observational study.World journal of emergency medicine · 2026
- Agreement and trending ability between non-invasive and invasive hemodynamic monitoring during cesarean section under spinal anesthesia: a prospective observational study.Frontiers in medicine · 2026
- Cuffless Devices for the Measurement of Blood Pressure: A Scientific Statement From the American Heart AssociationHypertension · 2025