Biosensors & bioelectronics

Non-invasive continuous interstitial fluid lactate monitoring via alternating-polarity reverse iontophoresis.

Wangwang Zhu, Mutian Wang, Jingyun Li, Jialing Wang, Wenjun Li, Dachao Li, Hao Zheng, Xingguo Zhang,

State Key Laboratory of Precision Measurement Technology and Instruments, Tianjin University, Tianjin, 300072, China. · State Key Laboratory of Precision Measurement Technology and Instruments, Tianjin University, Tianjin, 300072, China. Electronic address: zhenghao_0402@tju.edu.cn.

Key points

From the paper's conclusion
  • These results establish a viable route for non-invasive continuous ISF lactate monitoring, advancing wearable metabolic sensing toward practical applications.

Abstract

Non-invasive and continuous monitoring of interstitial fluid (ISF) lactate is critical for real-time metabolic assessment but remains unrealized. Reverse iontophoresis (RI) offers a promising route for transdermal ISF extraction; however, current fixed-polarity operation suffers from progressive electrode consumption and skin polarization, resulting in unstable extraction that hinders continuous monitoring. Here, we report a non-invasive and continuous ISF lactate monitoring (CLM) method based on alternating-polarity RI. A wearable CLM device integrating an enzymatic lactate sensor with paired RI electrodes is developed, in which the applied electric field is periodically reversed during extraction. This alternating-polarity operation enables cyclic electrode recovery while mitigating skin polarization, thereby stabilizing lactate extraction flux and supporting continuous measurement. In vitro studies using a Franz diffusion cell established a linear correlation of concentration between the extracted lactate and the subdermal one. This correlation enabled accurate prediction of the subdermal lactate later, yielding a fitted slope of 1.093, indicating strong agreement. In vivo validation during dynamic exercise further demonstrates that alternating-polarity RI tracks blood lactate levels more closely. The Pearson correlation coefficient between current response and blood lactate increased from negligible to strong correlation, with the mean value rising from 0.198 to 0.755. These results establish a viable route for non-invasive continuous ISF lactate monitoring, advancing wearable metabolic sensing toward practical applications.

Read the paper