The Journal of arthroplasty

Durability of a Proximally Coated, Dual-Tapered Cementless Femoral Stem Through 20 Years After Primary Total Hip Arthroplasty.

Nele Wagener, Thomas H Christensen, Thomas Bieganowski, David Novikov, Joseph D Zuckerman, Ran Schwarzkopf

Department of Orthopedic Surgery, NYU Langone Health, NYU Langone Orthopedic Hospital, New York, New York, USA. · Department of Orthopedic Surgery, NYU Langone Health, NYU Langone Orthopedic Hospital, New York, New York, USA. Electronic address: Ran.Schwarzkopf@nyulangone.org.

Observational · n=622

Key points

From the paper's conclusion
  • Under documented stem-specific institutional follow-up, this cementless femoral stem demonstrated high revision-free survival through 20 years. Late stem revisions were related to fracture or instability rather than aseptic loosening; no PJI-related stem revision was observed in institutional records. Most reinterventions involved non-femoral components.

Abstract

BACKGROUND: Studies of total hip arthroplasty (THA) through 20 years should distinguish failure of the femoral stem from revisions of components. This study evaluated revision-free survival of a proximally coated, dual-tapered cementless femoral stem and characterized causes of late revision and reoperation. METHODS: This retrospective single-center cohort included 730 THAs in 622 patients between February 1999 and June 2007. The primary endpoint was removal or exchange of the index stem. Revisions retaining the stem were analyzed separately. Kaplan-Meier survival was estimated from institutional follow-up. Hips without a qualifying postoperative orthopaedic visit or radiograph were censored at the index date and contributed no observation time. Stem-specific follow-up was available for 391 hips (53.6%), contributing 6,145.4 hip-years; median follow-up was 16.1 years (range, 3.9 to 26.3). Of these, 221 remained at risk at 15 years, including 112 at 20 years. RESULTS: There were three stems revised: two after periprosthetic fracture and one for instability/component malposition. There was no stem revision for aseptic loosening. There was one periprosthetic joint infection (PJI)-related head and liner exchange that was identified in institutional records and retained the index stem; no PJI-related stem revision was observed. Stem revision-free survival was 99.1% at 15 and 98.6% at 20 years. At 20 years, freedom from revision of any THA component and ipsilateral reoperation was 79.9 and 79.5%. Bearing wear/osteolysis accounted for 14 of 29 first component revisions (48.3%), due to conventional polyethylene. The 339 hips lacked qualifying postoperative institutional follow-up and were censored at the index date. Among patients represented by these hips, 17 were deceased; implant status remained unverified. CONCLUSION: Under documented stem-specific institutional follow-up, this cementless femoral stem demonstrated high revision-free survival through 20 years. Late stem revisions were related to fracture or instability rather than aseptic loosening; no PJI-related stem revision was observed in institutional records. Most reinterventions involved non-femoral components.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Durability of a Proximally Coated, Dual-Tapered Cementless Femoral Stem Through 20 Years After Primary Total Hip Arthroplasty — Science Skim