Effect of Prosthesis Selection on Gap Balancing in Robotic-Assisted Cruciate-Retaining and Posterior-Stabilized Total Knee Arthroplasty: A Prospective Randomized Controlled Trial.
Chao Yang, Kechao Zhu, Huiyong Dai, Dongyu Yang, Xianlong Zhang, Daoyun Chen, Qi Wang, Mengqi Cheng
Department of Orthopedics, Shanghai Jiao Tong University Affiliated Sixth People's Hospital, Shanghai 200233, China. · Department of Rehabilitation Medicine, Shanghai Jiao Tong University Affiliated Sixth People's Hospital, Shanghai 200233, China.
Key points
From the paper's conclusion- The CR design demonstrated superior balance in terms of the flexion-extension gap numerical value compared to the PS design. While the PS RA-TKA group required less posterior femoral bone resection and less external rotation of the femoral component, no significant differences in clinical outcomes were observed between the groups at a minimum of two years postoperatively. For most patients who had varus deformity and an intact posterior cruciate ligament (PCL), the CR implant design remains a reliable option. However, for those who have major flexion gap imbalance or PCL deficiency, the PS design is recommended.
Abstract
BACKGROUND: The study aimed to assess the disparities between robotic-assisted cruciate-retaining (CR) and posterior-stabilized (PS) total knee arthroplasty (RA-TKA) regarding flexion-extension gaps, bone resection amount, and postoperative knee function. METHODS: Prospectively randomized into two groups, CR RA-TKA (n = 60) and PS RA-TKA (n = 60), 120 patients were scheduled for RA-TKA. All patients underwent RA-TKA to execute adjusted implant positioning, flexion-extension gaps, bone resection amount, femoral component rotation, and alignment by using a robotic-assisted system, following functional alignment principles, between September 1, 2021, and December 31, 2022. Patients underwent clinical and radiographic evaluations at the 2-year postoperative follow-up. RESULTS: The extension gap results were similar between the two groups; the flexion gap was greater in the PS RA-TKA group, particularly on the lateral side (P = 0.027). Compared with the CR RA-TKA group, the PS RA-TKA group had significantly less bone resection of the lateral posterior condyle (P = 0.048). Although the PS group required less external rotation of the femoral component than the CR group, the difference was not statistically significant (P = 0.225). There were no significant differences in pre- or postoperative clinical scores between the two groups (P > 0.05). CONCLUSION: The CR design demonstrated superior balance in terms of the flexion-extension gap numerical value compared to the PS design. While the PS RA-TKA group required less posterior femoral bone resection and less external rotation of the femoral component, no significant differences in clinical outcomes were observed between the groups at a minimum of two years postoperatively. For most patients who had varus deformity and an intact posterior cruciate ligament (PCL), the CR implant design remains a reliable option. However, for those who have major flexion gap imbalance or PCL deficiency, the PS design is recommended.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparable clinical outcomes but different postoperative Dynamic Alignment of the Knee (DyAK) patterns between cruciate-retaining and posterior-stabilized inserts in functionally aligned robotic-assisted total knee arthroplasty.Journal of experimental orthopaedics · 2026
- Same patient, same total knee arthroplasty, different outcomes? A pilot study comparing conventional vs. OMNIBotics Robotic-Assisted HLS KneetecTM (Corin®).SICOT-J · 2026
- Cruciate-retaining versus condylar-stabilized reconstruction strategies in robotic-arm assisted total knee arthroplasty: a prospective, double-blinded, randomized controlled trial.Journal of robotic surgery · 2026
- Clinical outcomes of posterior-stabilized versus cruciate-retaining ANTHEM total knee arthroplasty using imageless robotic-assisted personalized alignment: a retrospective comparative cohort study.Journal of robotic surgery · 2026
- Robotic-assisted total knee arthroplasty: a narrative review of resection accuracy and individualized soft-tissue balance.Journal of robotic surgery · 2026
- Postoperative outcomes after conventional, robotic-arm-assisted, and pressure-sensor-guided cruciate-retaining total knee arthroplasty: a retrospective cohort study.Journal of robotic surgery · 2026