TDCR++ a code for radionuclide metrology in liquid scintillation and decay-data sensitivity analysis.
Benoît Sabot, Xavier Mougeot
Université Paris-Saclay, CEA, LIST, Laboratoire National Henri Becquerel (LNE-LNHB), Palaiseau, F-91120, France. Electronic address: benoit.sabot@cea.fr. · Université Paris-Saclay, CEA, LIST, Laboratoire National Henri Becquerel (LNE-LNHB), Palaiseau, F-91120, France.
Key points
From the paper's conclusion- The TDCR++ code provides batch processing of experimental datasets, traceable configuration and detailed output with a complete uncertainty budget to support radionuclide standardisation and decay data metrology studies.
Abstract
This work presents an analysis code dedicated to liquid scintillation counting based on the primary standard method known as the triple-to-double coincidence ratio (TDCR) method. The software combines evaluated decay data with event-based emission sampling to model multiple radionuclide transitions such as, α, β± and electron capture (EC) including complex schemes with internal conversion (IC), isomeric transitions and accompanying atomic-relaxation photon cascades. Deposited-energy components from charged particles (β, conversion and Auger electrons) and photons (γ-rays, X-rays, annihilation photons) are propagated to a scintillation-equivalent response through configurable quenching models. To eliminate biases induced by the non-linearity of quenching, TDCR++ implements a component-sum formulation in which individual energy-deposition components are treated separately. Atomic relaxation due to vacancies created by the EC and IC processes is modelled using corrected-EADL radiative and non-radiative transitions, with an automatic multi-vacancy merging strategy to preserve vacancy coupling within an event. The code also offers an prompt-versus-delayed treatment: the emission time of each de-excitation photon is sampled from an exponential distribution set by the half-life of the intermediate level, and compared with the coincidence window. The originality of this work relies in the uncertainty propagation which is performed by Monte Carlo sampling. The code is especially designed to perform analysis of variance studies in order to identify the predominant parameters in the calculation. It will help drive decay data studies in order to improve future measurements analysis. Originally developed in Python, the core of the code is implemented for CPU execution, with optional multi-core C++ and GPU (CYDA/CuPY) back ends for acceleration. However, the latter is not always the best option depending on the scheme compared to CPU/C++ solutions. The TDCR++ code provides batch processing of experimental datasets, traceable configuration and detailed output with a complete uncertainty budget to support radionuclide standardisation and decay data metrology studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Experimental validation of the scintiPulses shot-noise model for coincidence resolving-time dependencies in liquid scintillation counting.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Validation of the ray-tracing Monte-Carlo code tRAYcy for geometry and coincidence-summing corrections in environmental gamma-ray spectrometry.Journal of environmental radioactivity · 2026
- Monte Carlo uncertainty propagation in efficiency calculations for 4πγ counting of 134 Cs and 137 CsMetrologia · 2026
- Monte Carlo intercomparison of TG-43 dosimetric parameters for clinical192Ir and125I brachytherapy sources.Physics in medicine and biology · 2026
- Radionuclide Dependence of Distance-Dependent Spatial Resolution and Resolution Recovery in SPECT: A SIMIND Monte Carlo Simulation Study.Biomedical physics & engineering express · 2026
- Thermomechanical modelling and uncertainty reduction in cryogenic defined solid-angle alpha counting for primary Radon-222 standardisationMetrologia · 2026