Thermomechanical modelling and uncertainty reduction in cryogenic defined solid-angle alpha counting for primary Radon-222 standardisation
Behnam Khanbabaee, Anja Honig, Stefan Röttger
Key points
From the paper's conclusion- The results provide quantitative guidance for optimising cryogenic DSA source assemblies, while a complete uncertainty evaluation of the primary Radon-222 standardisation requires consideration of the actual spatial distribution of deposited activity and other measurement-model contributions.
Abstract
In defined solid-angle (DSA) alpha counting for the primary standardisation of Radon-222 activity, cryogenic operation can modify the measurement geometry through thermomechanical deformation of the source assembly and changes in the potential extent of the deposited source. Coupled heat-transfer and thermomechanical finite-element simulations were used to investigate the influence of nickel support rod length on the source–diaphragm distance and on the radial region in which Radon-222 condensation is thermodynamically possible. Thermomechanical deformation was found to be the dominant contribution to the uncertainty of the source–diaphragm distance for the investigated configurations. Shortening the support rod reduced this contribution but increased the thermally affected source radius. These effects were propagated to the geometry factor using a GUM-compliant uncertainty analysis. Under the assumed uncertainties of the investigated geometrical parameters, the calculated relative combined standard uncertainty associated with the geometry factor decreases from approximately 0.16 % for a 120 mm support rod to approximately 0.03 % for a 5 mm rod. The results provide quantitative guidance for optimising cryogenic DSA source assemblies, while a complete uncertainty evaluation of the primary Radon-222 standardisation requires consideration of the actual spatial distribution of deposited activity and other measurement-model contributions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Applicability of the defined solid angle method to the activity measurement of α-emitters for targeted radionuclide therapy.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Physics-based modelling of alpha detection in ZnS(Ag) scintillation cells for radon measurements.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Measurement of activity size distribution of radon and thoron progeny using DRPS/DTPS based cascade impactor.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Radon emanation coefficients of coal fly ash and soil: Measurement, wall flux geometry correction, and indoor dose implications.Journal of environmental radioactivity · 2026
- Investigation of key parameters influencing 222Rn exhalation rate from selected industrial and environmental samples using AlphaGUARD monitorScientific Reports · 2026
- Development of optimization-based algorithms for radiation source localization in nuclear emergenciesJournal of Radiation Research and Applied Sciences · 2026