Evaluation of a New CT Phantom Simulating the X-ray Attenuation of the Human Liver.

Zitong Cheng, Jian Geng, Zhe Guo, Yandong Liu, Yi Yuan, Bo Hu, Fengyun Zhou, Jing Zheng,

Department of Radiology, Beijing Hospital, Capital Medical University, National Center for Orthopaedics, Beijing, China (Z.C., J.G., Z.G., Y.L., Y.Y., B.H., J.Z., Q.Z., D.Y., L.W., X.C.). · Department of Radiology, Chongqing University Cancer Hospital, Chongqing, China (F.Z.).

Key points

From the paper's conclusion
  • DLCT measurements of Zeff and ED values in human livers were not consistent with similar scans of a phantom designed to simulate liver composition according to ICRU Report 46. However, measurements of the 100% AT insert were broadly consistent with the expected relationship between white AT and pure fat.

Abstract

RATIONALE AND OBJECTIVES: To evaluate a new computed tomography (CT) liver phantom based on the International Commission on Radiation Units (ICRU) Report 46 composition of liver and adipose tissue (AT) and compare it with human studies calibrated against magnetic resonance imaging (MRI) proton density fat fraction (PDFF). MATERIALS AND METHODS: A liver phantom with seven inserts, ranging from 100% lean liver to 100% AT, was commissioned to reproduce the X-ray attenuation of the human liver by simulating the effective atomic number (Zeff) and electron density (ED) set out in ICRU Report 46. Five 120 kVp scans were acquired on each of nine CT scanners across five Chinese cities. Liver fat was measured using Mindways software calibrated to reproduce MRI PDFF measurements. Phantom data were compared with human CT and PDFF studies to assess reliability. RESULTS: Phantom insert measurements showed that 0% AT corresponded to 10.0% liver fat by Mindways, while 100% AT corresponded to 72.3%. Comparison of Mindways measurements with liver PDFF in 400 healthy volunteers showed agreement between the two fat scales. Dual-layer CT (DLCT) measurements of Zeff and ED matched the phantom manufacturer's specifications for the 0% AT insert (Zeff = 7.41; ED = 105.2%). However, linear regression of Zeff and ED against MRI PDFF in 99 healthy volunteers gave intercepts of Zeff = 7.18 and ED = 107.1% at 0% PDFF. CONCLUSION: DLCT measurements of Zeff and ED values in human livers were not consistent with similar scans of a phantom designed to simulate liver composition according to ICRU Report 46. However, measurements of the 100% AT insert were broadly consistent with the expected relationship between white AT and pure fat.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers