Monte Carlo investigation of anode-filter combinations on X-ray spectra in conventional radiology using PHITS.

Alex Eclador Ngankem, Joel Cebastien Guembou Shouop, Francis Chamberlain Djoumessi Zamo, Daniel Bongue, Maurice Moyo Ndontchueng

Pole de Recherche, d'Innovation et d'Entrepreneuriat (PRIE), Institut Universitaire de la Côte, P.O. Box: 3001, Douala, Cameroun. ngankem.eclador@myiuc.com. · Radiological Safety and Nuclear Security Authority, P.O. Box: 33732, Yaoundé, Cameroon. sebastianguembou@gmail.com.

Key points

From the paper's conclusion
  • The results indicate that Tungsten (W) and Gold (Au) anodes (K-lines: 59-68 keV) are optimal for radiography, whereas Molybdenum (Mo) and Rhodium (Rh) anodes (K-lines: 17-23 keV) are superior for mammography.

Abstract

This research project investigates conventional radiology protocols by using PHITS (Particle and Heavy Ion Transport code System) Monte Carlo simulations to evaluate anode-filter configurations at tube voltages up to 150 kV. The study's innovation lies in its quantitative comparison of simulated spectra with SpekCalc, showing a mean deviation of less than 5%. The results indicate that Tungsten (W) and Gold (Au) anodes (K-lines: 59-68 keV) are optimal for radiography, whereas Molybdenum (Mo) and Rhodium (Rh) anodes (K-lines: 17-23 keV) are superior for mammography. By quantifying the impact of heavy filtration on beam hardening, this work establishes a robust framework for improving image contrast and supporting dose optimisation through appropriate selection of tube parameters.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers