Characterization of open-source BSREM reconstruction for 123I-FP-CIT spect on a conventional anger-type SPECT/CT system: a phantom study.
Takuro Shiiba, Shingo Tanahashi, Kohei Takasu, Yutaro Hamajima, Kosuke Yamashita, Masakazu Tsujimoto
Department of Medical Image Analysis, Division of Advanced Biomedical Sciences, Faculty of Life Sciences, Kumamoto University, 4-24-1, Kuhonji, Chuo-ku, Kumamoto , Kumamoto, 862-0976, Japan. shiiba@kumamoto-u.ac.jp. · Department of Radiology, Fujita Health University Hospital, 1-98, Dengakubo, Kutsukake-cho, Toyoake, Aichi, 470-1192, Japan.
Key points
From the paper's conclusion- This study characterized the recovery-variability trade-off, count-level dependence, and region-specific behavior of open-source block-sequential regularized expectation maximization (BSREM) reconstruction for 123I-FP-CIT single-photon emission computed tomography acquired on a conventional anger-type SPECT/CT system.
Abstract
This study characterized the recovery-variability trade-off, count-level dependence, and region-specific behavior of open-source block-sequential regularized expectation maximization (BSREM) reconstruction for 123I-FP-CIT single-photon emission computed tomography acquired on a conventional anger-type SPECT/CT system. Striatal phantom data with normal and asymmetric reduced-putaminal uptake were reconstructed using literature-based ordered-subsets expectation maximization (OSEM) implemented in the vendor software, PyTomography OSEM, and PyTomography BSREM. A BSREM operating point (β = 0.08, γ = 6.0) was selected using background coefficient of variation (CV) and whole-striatal uptake ratio (SUR). Reduced-count datasets at 75%, 50%, and 25% of the original counts were generated by binomial thinning. Metrics included CV, regional SUR, asymmetry index, caudate-to-putamen ratio (CPR), and full width at half maximum (FWHM). In the normal phantom, BSREM yielded lower CV and higher SUR than both OSEM implementations. In the abnormal phantom, BSREM generally yielded higher regional SURs, whereas mean caudate FWHM was 22.6 mm for vendor-software OSEM and 16.9 mm for PyTomography OSEM, and 17.5 mm for BSREM. However, reduced left-putaminal uptake remained underestimated, and putaminal asymmetry moved further from their theoretical values. At the 25% count level, background CV increased to 30.0 ± 6.1% with BSREM and 28.0 ± 5.9% with PyTomography OSEM. Open-source BSREM therefore showed task-, region-, and count-dependent behavior, supporting task- and count-specific parameter selection before clinical application.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optimization of spectral photon-counting CT reconstruction parameters for improved detection of peritoneal metastases compared with energy-integrating detector dual-energy CT: a phantom study.European radiology experimental · 2026
- Toward an optimized PET reconstruction protocol for spinal cord: superiority of OSEM with TOF and PSF.EJNMMI physics · 2026
- Proof-of-Concept Study of Task-Based Adaptive SPECT Imaging via System Sampling Strategies on a 12-Head Robotic SPECT Platform.IEEE transactions on medical imaging · 2026
- Kinematic Total-Body PET Video Reconstruction via Hierarchical List-Mode EM and Validation-Adaptive Space-Time RegularisationarXiv · 2026
- Phantom evaluation of dynamic SPECT imaging with a ring-type semiconductor SPECT/CT system: comparison with an anger-type systemEJNMMI Physics · 2026
- Quantitative Reconstructions for Renal Sestamibi SPECT/CT: A Modeled Comparison of 360-Degree Versus 180-Degree Acquisitions.Nuklearmedizin. Nuclear medicine · 2026