A method for distance accuracy evaluation on the reformatted oblique plane of CT images.
Choirul Anam, Nafisah N 'Alamah, Evi Setiawati, Ariij Naufal, Riska Amilia, Heri Sutanto, Eko Hidayanto, Lydia P W S Kuntjoro,
Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang, 50275, Central Java, Indonesia. anam@fisika.fsm.undip.ac.id. · Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Tembalang, Semarang, 50275, Central Java, Indonesia.
Key points
From the paper's conclusion- The proposed method for evaluating distance accuracy on reformatted oblique images using the ACR CT phantom is easily implemented.
Abstract
This study proposes a new method for evaluating distance accuracy on the reformatted oblique plane using the American College of Radiology (ACR) computed tomography (CT) phantom. Evaluations were performed on oblique images reformatted from axial, sagittal, and coronal images. Oblique images from axial images were reformatted through two radiopaque markers commonly used for distance accuracy on the axial image. Oblique images from sagittal and coronal images were reformatted through eight radiopaque markers at the edge of the phantom usually used to perform centering of the phantom in the x, y, and z directions. Axial images were scanned with two pitches of 0.86 and 1.0. Oblique images were reformatted using IndoQCT and RadiAnt. The distance accuracy was evaluated on the reformatted oblique plane using an electronic caliper available in these two platforms. The percentage differences of measured distances on the reformatted oblique image using IndoQCT were within 2% of the actual distances. The average percentage differences for a pitch of 0.86 was 0.78%, and for pitch of 1.0, it was 0.57%. The percentage differences of measured distances using RadiAnt were within 1% of the actual distances. The average percentage differences for a pitch of 0.86 was 0.22%, and for a pitch of 1.0, it was 0.29%. The proposed method for evaluating distance accuracy on reformatted oblique images using the ACR CT phantom is easily implemented. We found that the percentage difference of distances from the IndoQCT and RadiAnt was within 2%. It is recommended that distance accuracy on the reformatted oblique plane is performed in regular quality control (QC).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Elliptical acquisition provides measurement accuracy comparable to CT in an experimental model of hip surgery.Orthopaedics & traumatology, surgery & research : OTSR · 2026
- An exploratory study on the influence of CT slice spacing on the positioning accuracy of 3D-printed patient-specific instruments designed for a distal radius osteotomy.Journal of orthopaedic surgery and research · 2026
- Impact of reconstruction-based misalignment correction on image quality, reader agreement, and diagnostic confidence in single-source photon-counting CT.European journal of radiology · 2026
- Visually detectable pointer deviation of the nextar surgical navigation system with c-arm, o-arm and ct imaging: a bench-top study using a two-dimensional phantom.BMC musculoskeletal disorders · 2026
- Automatic correction of propagation distance in ptychographic image reconstruction.Journal of the Optical Society of America. A, Optics, image science, and vision · 2026
- Localiser-Based Software-Assisted Patient Positioning in Chest CT: A Propensity Score-Matched Study.Diagnostics (Basel, Switzerland) · 2026