Outcomes of Modified Microvascular Plugs as a Pulmonary Flow Restrictors in Congenital Heart Disease.
Brian Tran, Jacob Shalkevich, Joseph Graziano, Byron Garn, Daniel A Velez, Mohamad Alaeddine, Arash Sabati, Saleem Almasarweh
Phoenix Children's Hospital, Cardiology, Phoenix, AZ, USA. btran2@phoenixchildrens.com. · Phoenix Children's Hospital, Pediatrics, Phoenix, AZ, USA.
Key points
From the paper's conclusion- Manually modified microvascular plugs (M-MVPs) have emerged as transcatheter pulmonary flow restrictors (PFRs) in neonates and infants with complex congenital heart disease and pulmonary overcirculation.
Abstract
Manually modified microvascular plugs (M-MVPs) have emerged as transcatheter pulmonary flow restrictors (PFRs) in neonates and infants with complex congenital heart disease and pulmonary overcirculation. Standardized device modification methods and short-term outcomes remain limited. We performed a retrospective single-center clinical data review of consecutive patients who underwent transcatheter pulmonary flow restriction using M-MVPs from January 2023 through June 2025. Data collected included patient characteristics, procedural details, peri-procedural complications, surgical outcomes, device explantation, and mortality. Twenty-one neonates and infants (median age: 16 days (IQR 8-22); median weight: 2.8 kg (IQR 2.3-3.9) received 44 M-MVP devices. Femoral venous access was used in 20/21 cases, with carotid access required in one patient. Median angiographic branch pulmonary artery diameters were 5.1 mm (left) and 5.3 mm (right). Fifteen patients underwent surgical explantation at a median 43 days (IQR 37.5-58.5 days), with one patient supported for 236 days. All devices were surgically explanted without major intraoperative complications or device related stenosis requiring pulmonary arterioplasty. Three patients required additional device implantation for distal migration or incomplete pulmonary artery coverage, and two patients ultimately required conversion to surgical main pulmonary artery banding. Interval mortality was 29% (6/21) in a cohort with substantial comorbidity, physiologic instability, and limited surgical candidacy. Exploratory physiologic analysis demonstrated higher renal near-infrared spectroscopy values among patients who survived to surgery, although physiologic responses following implantation were variable. In this retrospective descriptive experience, M-MVPs provided technically feasible temporary pulmonary flow restriction in carefully selected high-risk neonates and infants. Given the small sample size, heterogeneous cohort, and lack of a control group, the findings presented should be considered descriptive and hypothesis generating.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Transcatheter pulmonary flow restriction in congenital heart disease: evolving concepts and future perspectives.Future cardiology · 2026
- In Silico Insights into Device-Artery Interactions in Endovascular Pulmonary Flow Restrictors for Neonates with Congenital Heart Defects.Cardiovascular engineering and technology · 2026
- Two slits good, four slits bad? Assessing the hemodynamic impact of modifications to microvascular plugs in pulmonary flow restriction using a bench-top model.Cardiology in the young · 2026
- Immediate Hemodynamic Effects of Pulmonary Artery Flow Restrictors in Univentricular Circulation: Hypothesis-Generating Insights from High-Fidelity Physiologic Streaming Data Using Machine-Learning.Pediatric cardiology · 2026
- Transcatheter atrial flow regulator implantation in children with pulmonary arterial hypertension in resource-limited settings: a single-centre experience.Cardiology in the young · 2026
- Transcatheter Closure of Residual Shunt Using the Coil-in-Plug Method After Amplatzer Vascular Plug Implantation for Tubular Patent Ductus Arteriosus in Infants: Two Case Reports.Pediatric cardiology · 2026