Bioengineering (Basel, Switzerland)

In Vitro Hemodynamic Assessment of a Novel Pulmonary Visceral Pleura (PVP) Transcatheter Aortic Valve Replacement Prototype Based on Optical and Acoustic Measurements.

Colten Alme, Santu Golder, Xiao Lu, Haoxiang Luo, Yan Zhang

Department of Mechanical Engineering, North Dakota State University, Fargo, ND 58105, USA. · California Medical Innovations Institute (CalMI2), San Diego, CA 92121, USA.

Preclinical

Key points

From the paper's conclusion
  • The silicone valve opened to 1.1 cm2 with a gradient of 36.9 mmHg, consistent with moderate-to-severe stenosis.

Abstract

Structural valve degradation remains a key limitation of bioprosthetic transcatheter aortic valve replacement (TAVR) devices, whose bovine or porcine pericardium leaflets are prone to calcification and long-term failure. Pulmonary visceral pleura (PVP), a thin, elastin-rich tissue, has emerged as a promising alternative leaflet material. This study presents an in vitro hemodynamic evaluation of a prototype PVP valve under physiologically representative pulsatile flow in a custom benchtop flow loop. A multimodal framework combined four techniques: high-speed imaging with U-Net deep learning segmentation of the geometric orifice area, quantitative Doppler velocity extraction developed on a Verasonics research ultrasound platform, double-pulse particle image velocimetry (PIV) of the downstream flow field, and transvalvular pressure measurement. The PVP valve was compared with a stiff silicone valve representing a stenotic reference condition. At 75 bpm and 5 L/min, the PVP valve reached a peak geometric orifice area of 2.91 cm2, a mean pressure gradient of 4.3 mmHg, and a peak jet velocity of 2.46 m/s, meeting the ASE velocity and gradient criteria for normal prosthetic valve function. The silicone valve opened to 1.1 cm2 with a gradient of 36.9 mmHg, consistent with moderate-to-severe stenosis. The framework provides a basis for characterizing new bioprosthetic valve materials and benchmark data for fluid-structure interaction modeling.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

In Vitro Hemodynamic Assessment of a Novel Pulmonary Visceral Pleura (PVP) Transcatheter Aortic Valve Replacement Prototype Based on Optical and Acoustic Measurements — Science Skim