PHITS-based validation of occupational doses during coronary angiography procedures.
Marianna Lourdes Marie L Grande, Ranzivelle Marianne Roxas-Villanueva
Department of Science and Technology-Philippine Nuclear Research Institute, Commonwealth Avenue, Diliman, Quezon City, Philippines. annamariegrande@gmail.com. · Institute of Physics, University of the Philippines Los Baños, Laguna, Philippines.
Key points
From the paper's conclusion- These findings validate the PHITS model as a reliable dosimetric tool under the simulated conditions and demonstrate the influence of gantry angulation on scatter field geometry.
Abstract
Occupational radiation monitoring in interventional cardiology remains inconsistent in many resource-limited settings, where personal dosimetry programs are incomplete or absent. This study employed the Particle and Heavy Ion Transport code System (PHITS) to estimate occupational radiation doses to medical staff across nine standard projection angulations during diagnostic coronary angiography. The simulation modeled a point x-ray source at 65 kVp with a water phantom representing the patient, and used T-point tallies to compute ambient dose equivalent H*(10) at staff positions corresponding to the cardiologist, fellow, radiation technologist, and nurse. Results show the cardiologist consistently received the highest dose across all projections, with the AP/AP and LAO10/CRAN30 angulations producing peak values and RAO15/CRAN45 the lowest. The cumulative per-procedure cardiologist dose falls within the range reported in published international literature and agrees to within approximately 10% of a locally established Philippine benchmark. A pronounced dose gradient was observed across staff roles, with the cardiologist receiving substantially higher exposure than other personnel. These findings validate the PHITS model as a reliable dosimetric tool under the simulated conditions and demonstrate the influence of gantry angulation on scatter field geometry. The study supports the use of Monte Carlo-based computational dosimetry as a practical alternative for occupational dose assessment in settings lacking consistent personal monitoring, with direct implications for radiation protection planning in Philippine catheterization laboratories.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Eye lens dose in interventional cardiology personnel in Vietnam: direct Hp(3) measurement, indirect method comparison, and fluoroscopy-time-based dose estimation.Radiation protection dosimetry · 2026
- Occupational radiation exposure and radiation protection practices in interventional radiology and cardiology: A multicenter study in FranceDiagnostic and Interventional Imaging · 2026
- From whole-body dosimetry to local dose assessment: a 3-y experience in interventional practice.Radiation protection dosimetry · 2026
- Real-time occupational radiation dosimetry in anesthesiology personnel during fluoroscopy-guided interventions: a single-center descriptive and exploratory study.Frontiers in public health · 2026
- Identifying High-Dose Views and Modes in the Catheterization Laboratory: A Run-Level Mixed-Effects Analysis.Turk Kardiyoloji Dernegi arsivi : Turk Kardiyoloji Derneginin yayin organidir · 2026
- Quantifying per-event dose-equivalent rates of echocardiographers and interventional cardiologists during structural heart interventions.Radiological physics and technology · 2026