Device-procedure attribution of reported harm in robotic-assisted bronchoscopy: a MAUDE study informing safety surveillance of robotic surgical systems.
Zitian Wang, Baishun Liu, Cailing Wei, Xiaohui Chen, Yuwan Gao, Qi Xu, Juan Wang
School of Future Technology, Xi'an Jiaotong University, Xi'an, China. · School of Medicine, Xi'an Jiaotong University, Xi'an, China.
Key points
From the paper's conclusion- We evaluated narrative-supported attribution in US Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) reports involving the Ion and Monarch platforms for robotic-assisted bronchoscopy, a navigation-dependent minimally invasive thoracic intervention.
Abstract
Postmarket surveillance of robotic surgical systems requires distinguishing harm associated with platform performance from harm arising through procedural, anatomical, anesthetic, or perioperative pathways. We evaluated narrative-supported attribution in US Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) reports involving the Ion and Monarch platforms for robotic-assisted bronchoscopy, a navigation-dependent minimally invasive thoracic intervention. Death- and injury-designated reports received through 31 May 2026 were reviewed using a prespecified framework. Two independent reviewers assessed whether patient harm was documented and classified eligible reports as device-related, procedure-related, mixed device-procedure, or indeterminate; disagreements were adjudicated by a third reviewer. Among 1,691 reports, 1,638 (96.9%) contained narrative evidence of harm. Among harm-eligible reports, narrative-supported attribution was procedure-related in 1,429 (87.2%), device-related in 27 (1.6%), mixed device-procedure in 31 (1.9%), and indeterminate in 151 (9.2%). Harm-eligibility agreement was 99.7% (κ = 0.952), whereas attribution agreement among reports judged eligible by both reviewers was 87.8% (κ = 0.587). Indeterminate attribution was more frequent among death- than injury-designated reports (34.5% vs. 7.4%); no death-designated report met the prespecified criteria for device-related attribution. Submitted narratives therefore more often supported procedural or perioperative mechanisms than explicit malfunction-to-harm pathways, but this finding neither establishes operator error nor excludes undocumented device contribution. Mechanism-based attribution may improve triage for engineering investigation, procedural quality review, and evidence preservation across robotic surgical systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Injury labels and patient-problem codes in FDA adverse event reports for robotic arthroplasty platforms: a cross-sectional analysis of 12,175 MAUDE reports.Journal of robotic surgery · 2026
- Laparoscopic and Robotic Device Malfunction: A Scoping Review.Journal of patient safety · 2026
- Perioperative adverse events and device malfunctions associated with ROSA cranial stereotactic robotic surgery a descriptive analysis of FDA MAUDE reports from 2016 to 2026.Journal of robotic surgery · 2026
- Post-market safety signals of energy-based surgical devices: a cross-specialty disproportionality analysis of the FDA MAUDE database with robotic-platform validation and regulatory triangulation, 2015-2025.Journal of robotic surgery · 2026
- Device-related adverse event reports in robotic hernia repair: a disproportionality analysis of the FDA MAUDE database.Journal of robotic surgery · 2026
- Federated Bayesian Surveillance of Mechanical Thrombectomy Adverse Events: A Population Risk Layer for Surgical Digital TwinsarXiv · 2026