Journal of robotic surgery

Device-related adverse event reports in robotic hernia repair: a disproportionality analysis of the FDA MAUDE database.

Yiping Chen, Yuhai Hu, Chunzhong Lin

Department of Hepatopancreatobiliary Surgery, the First Affiliated Hospital, Fujian Medical University, Fuzhou, 350005, China. cypgg@fjmu.edu.cn. · Department of Hepatopancreatobiliary Surgery, the First Affiliated Hospital, Fujian Medical University, Fuzhou, 350005, China.

Key points

From the paper's conclusion
  • We analyzed 3,686 hernia repair and 3,010 cholecystectomy events.

Abstract

Robotic hernia repair has spread without demonstrated superiority, and its device-related adverse event reports are unstudied. We compared them with reports from other robotic procedures, prespecifying cholecystectomy as the primary comparator. Using complete US Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience files (cutoff 8 October 2026), we identified operations by prespecified narrative rules and merged 2022-2026 reports into events. Reporting odds ratios (ROR) with 95% confidence intervals (CI) covered 18 reporter-coded outcomes, three of them hypotheses: jaw entrapment (clip appliers excluded), fragmentation or detachment, and tissue injury. Sensitivity and post hoc analysis varied comparator, device category and event merging; two surgeons double-coded 50 reports. We analyzed 3,686 hernia repair and 3,010 cholecystectomy events. Hernia repair reports more often described entrapment (7.6% vs. 1.9%; ROR 4.19, 95% CI 2.87-6.10), fragmentation (17.0% vs. 10.4%; 1.75, 1.52-2.03) and tissue injury (5.1% vs. 2.8%; 1.87, 1.44-2.43). Against colorectal surgery, none was a signal. Within device categories, entrapment remained elevated (pooled ROR 2.44, 1.64-3.65), whereas fragmentation reversed (0.61, 0.48-0.77) and tissue injury varied by category (1.23, 0.92-1.64). Estimates were stable across event-merging rules. All 34 sampled cohort reports were confirmed; estimated cohort sensitivity was 94.0%. Relative to cholecystectomy, robotic hernia repair reports were enriched for jaw entrapment, which persisted within device categories but not against colorectal surgery; fragmentation and tissue-injury signals depended on device category. These hypothesis-generating findings describe report composition, not incidence, clinical risk or device causation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers