Device-related adverse event reports in robotic hernia repair: a disproportionality analysis of the FDA MAUDE database.
Yiping Chen, Yuhai Hu, Chunzhong Lin
Department of Hepatopancreatobiliary Surgery, the First Affiliated Hospital, Fujian Medical University, Fuzhou, 350005, China. cypgg@fjmu.edu.cn. · Department of Hepatopancreatobiliary Surgery, the First Affiliated Hospital, Fujian Medical University, Fuzhou, 350005, China.
Key points
From the paper's conclusion- We analyzed 3,686 hernia repair and 3,010 cholecystectomy events.
Abstract
Robotic hernia repair has spread without demonstrated superiority, and its device-related adverse event reports are unstudied. We compared them with reports from other robotic procedures, prespecifying cholecystectomy as the primary comparator. Using complete US Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience files (cutoff 8 October 2026), we identified operations by prespecified narrative rules and merged 2022-2026 reports into events. Reporting odds ratios (ROR) with 95% confidence intervals (CI) covered 18 reporter-coded outcomes, three of them hypotheses: jaw entrapment (clip appliers excluded), fragmentation or detachment, and tissue injury. Sensitivity and post hoc analysis varied comparator, device category and event merging; two surgeons double-coded 50 reports. We analyzed 3,686 hernia repair and 3,010 cholecystectomy events. Hernia repair reports more often described entrapment (7.6% vs. 1.9%; ROR 4.19, 95% CI 2.87-6.10), fragmentation (17.0% vs. 10.4%; 1.75, 1.52-2.03) and tissue injury (5.1% vs. 2.8%; 1.87, 1.44-2.43). Against colorectal surgery, none was a signal. Within device categories, entrapment remained elevated (pooled ROR 2.44, 1.64-3.65), whereas fragmentation reversed (0.61, 0.48-0.77) and tissue injury varied by category (1.23, 0.92-1.64). Estimates were stable across event-merging rules. All 34 sampled cohort reports were confirmed; estimated cohort sensitivity was 94.0%. Relative to cholecystectomy, robotic hernia repair reports were enriched for jaw entrapment, which persisted within device categories but not against colorectal surgery; fragmentation and tissue-injury signals depended on device category. These hypothesis-generating findings describe report composition, not incidence, clinical risk or device causation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Reported device problems and patient outcomes in da Vinci robotic robotic gastric and upper-gastrointestinal procedures: an FDA MAUDE database analysis.Journal of robotic surgery · 2026
- Hybrid robotic transversus abdominis release (hrTAR) for complex ventral hernias: a systematic review and preliminary synthesis.Journal of robotic surgery · 2026
- Injury labels and patient-problem codes in FDA adverse event reports for robotic arthroplasty platforms: a cross-sectional analysis of 12,175 MAUDE reports.Journal of robotic surgery · 2026
- Laparoscopic and Robotic Device Malfunction: A Scoping Review.Journal of patient safety · 2026
- Robotic-Assisted Versus Open Transversus Abdominis Release (TAR) for Complex Ventral Hernia Repair: A Systematic Review and Meta-Analysis.Cureus · 2026
- Comparative outcomes of robotic versus laparoscopic inguinal hernia repair: a systematic review and meta-analysisJournal of Robotic Surgery · 2026