Precise and minimally invasive localization of pelvic nodes during robot-assisted surgery using image-guided navigation.
Laura Aguilera Saiz, Wouter J Heerink, Harald C Groen, Marijn A J Hiep, Lisanne L P G Venix, Enzo Kerkhof, Henk G van der Poel, Esther M K Wit,
Department of Surgical Oncology, Netherlands Cancer Institute, Amsterdam, The Netherlands. l.aguilera@nki.nl. · Department of Surgical Oncology, Netherlands Cancer Institute, Amsterdam, The Netherlands.
Key points
From the paper's conclusion- Integrating ultrasound registration, instrument tracking and enhanced visualization during robot-assisted surgery has the potential to improve clinical use and localization of navigation in robot-assisted surgeries. These techniques allow to safely identify and remove pre-defined small pelvic lymph nodes in the pelvis.
Abstract
PURPOSE: Image-guided surgical navigation can enhance surgical precision and safety. The rise of robot-assisted surgeries has increased the need of incorporating navigation systems. Our goal is to evaluate the integration of three novel techniques on registration, instrument tracking and visualization with navigation using the robot-assisted sentinel lymph node biopsy (SLNB). METHODS: A prospective feasibility study was performed at the Netherlands Cancer Institute (November 2023-July 2025). In total, 30 patients scheduled for SLNB participated. First, registration and localization accuracy were analyzed. Secondly, surgeons answered questionnaires to evaluate the ultrasound registration, instrument tracking and enhanced visualization. Virtual reality (VR) and augmented reality (AR) were compared in the last ten patients. Feasibility was determined with the sentinel nodes (SN) percentage that were successfully localized by navigation and validated ex vivo with the gamma probe. RESULTS: Per-protocol, the first ten patients were used for workflow optimization; the subsequent 20 patients were included in the analysis. Ultrasound registration was fast (7 min), accurate (0.1 cm) and intuitive. Tracked instruments were well-integrated and helped to localize precisely the SN (0.3 cm). Surgeons preferred VR while opening the resection path to find the SN and AR in proximity for exact localization. Using navigation, successful identification was achieved in 90% (45/50) SN. No navigation-related complications occurred. CONCLUSIONS: Integrating ultrasound registration, instrument tracking and enhanced visualization during robot-assisted surgery has the potential to improve clinical use and localization of navigation in robot-assisted surgeries. These techniques allow to safely identify and remove pre-defined small pelvic lymph nodes in the pelvis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Intraoperative support using 3D reconstructed images for sentinel lymph node biopsy in endometrial cancer surgeryAsian Journal of Surgery · 2026
- Trackerless Augmented Reality Navigation for Robotic Partial Nephrectomy using Simultaneous Localization and Mapping.Research square · 2026
- Image-guided and augmented reality-assisted robotic surgery.Current opinion in urology · 2026
- Multimodal Image-based Robotic Assistance and Guidance Environment (MIRAGE): A Novel Unified Framework for Surgical Navigation and VisualizationarXiv · 2026
- Visually detectable pointer deviation of the nextar surgical navigation system with c-arm, o-arm and ct imaging: a bench-top study using a two-dimensional phantom.BMC musculoskeletal disorders · 2026
- Image-guided navigation with electromagnetic tracking during robot-assisted partial nephrectomy.Surgical endoscopy · 2026