Freehand 3D ultrasound imaging of the femoral head and neck.
Andrew D Michels, Orion P C Wiersma, Grace Yu, Aratha Thanamayooran, Ivan Wong, Robert B A Adamson
School of Biomedical Engineering, Faculty of Medicine, Dalhousie University, Halifax, Canada. · Division of Orthopaedic Surgery, Department of Surgery, Nova Scotia Health, Halifax, Canada.
Key points
From the paper's conclusion- The combination of 3D optical tracking, ultrasound imaging, and deep learning-based segmentation using models trained on public datasets and enhanced by transfer learning can produce bone surface maps of the femoral head with sufficient accuracy to resolve clinically significant femoral head deformities. The system is suitable for clinical deployment.
Abstract
PURPOSE: To evaluate the performance of a novel freehand 3D ultrasound system for assessing morphological deformities of the femoral head and neck associated with femoroacetabular impingement syndrome (FAIS). METHODS: The system integrates an off-the-shelf optical tracker with a low-cost point-of-care ultrasound probe. Bone surfaces were segmented from B-mode images using a deep learning feature pyramid network trained on a publicly available dataset and refined via transfer learning with curated, manually segmented images. Deterministic algorithms were used to generate bone surface point clouds and meshes. System accuracy was assessed by comparing reconstructed bone surfaces to computed tomography (CT) in four cadaveric hips and two patient hips. The system's feasibility for clinical imaging was evaluated in five healthy volunteers. RESULTS: The ultrasound reconstructions achieved geometric agreement with CT, within-subject repeatability and within-subject reproducibility of 0.4-0.6 mm. Alpha angles derived from CT and ultrasound scan agreed to within 0.6°. Given that clinically significant bone deformities associated with FAIS are ≥ 2 mm and clinically significant differences in alpha angle are > 5°, the system's performance appears adequate for clinical morphometry. In vivo imaging demonstrated feasibility for clinical use. CONCLUSION: The combination of 3D optical tracking, ultrasound imaging, and deep learning-based segmentation using models trained on public datasets and enhanced by transfer learning can produce bone surface maps of the femoral head with sufficient accuracy to resolve clinically significant femoral head deformities. The system is suitable for clinical deployment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- TP4.4 A Three-Dimensional Morphological Analysis of Cam Lesions in Femoroacetabular Impingement Syndrome Yields Three Distinct Phenotypes: Moving Beyond the Alpha AngleJournal of Hip Preservation Surgery · 2026
- Algorithmic determination of femoral neck axis in Birmingham hip resurfacing: A parametric CAD-based method.Acta of bioengineering and biomechanics · 2026
- Preoperative measurement of femoral head using 3D-CT spherical fitting technique for the guidance of femoroplasty in arthroscopic treatment of cam-type femoroacetabular impingement: a cohort validation.Frontiers in medicine · 2026
- Automated Femoral, Acetabular, and Global Offset Measurements on Pelvis Radiograph Using Deep Learning.Hip & pelvis · 2026
- FP7.1 A pilot study of AI-driven 3D reconstruction from a single radiograph for acetabular version assessmentJournal of Hip Preservation Surgery · 2026
- EP4.11 Accurate CT Segmentation and Reliable 3D Morphometric Measurements to Support Clinical Interpretation in FAIJournal of Hip Preservation Surgery · 2026