Algorithmic determination of femoral neck axis in Birmingham hip resurfacing: A parametric CAD-based method.
Michał P Kowalik, Dawid Knaś, Krzysztof Zawiśliński
Institute of Aeronautics and Applied Mechanics (IAAM), Warsaw University of Technology, Warsaw, Poland.
Key points
From the paper's conclusion- The method provides reproducible, automated femoral neck axis determination with plus/minus 0.5 degrees repeatability - approximately three times smaller than CT-induced uncertainty and an order of magnitude below the 5-7 degree errors of conventional jigs. Future work should validate the method against clinical outcomes in larger cohorts.
Abstract
PURPOSE: Birmingham hip resurfacing requires precise determination of the femoral head-neck axis to minimize risks of neck fracture, cortical notching, and component malposition. Existing methods rely on intraoperative estimation, generic guides, or semi-manual planning, all susceptible to anatomical variability. This study presents an automated parametric algorithm for determining the femoral neck axis from computed tomography (CT)-derived models to support patient-specific surgical planning. METHODS: The method utilizes segmented CT data converted to Standard for the Exchange of Product model data format. Following artifact removal, the algorithm performs iterative cross-sectional analysis along the neck axis, identifying characteristic anatomical points: the neck isthmus, head centroid, and head apex. The axis is approximated from these points and refined through rotational optimization in 5 degree and 0.5 degree increments. The algorithm was evaluated on 4 femoral models in 5 spatial orientations each (20 test cases). RESULTS: The algorithm achieved a mean validation score of 4.1/5.0 (82%). Three models with high-quality meshes scored 4-5, while one model with extensive artifacts scored 1-2, confirming sensitivity to input data quality. Repeatability was plus/minus 0.5 degrees in angular orientation and plus/minus 0.2 mm in isthmus position. Error propagation analysis indicated that CT-induced uncertainty (approximately 1.6 degrees upper bound) constitutes the limiting factor, not algorithmic precision. Physical prototyping confirmed viability for patient-specific instrument design. CONCLUSION: The method provides reproducible, automated femoral neck axis determination with plus/minus 0.5 degrees repeatability - approximately three times smaller than CT-induced uncertainty and an order of magnitude below the 5-7 degree errors of conventional jigs. Future work should validate the method against clinical outcomes in larger cohorts.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effects of variable bolt position of femoral neck system on Pauwels III fractures with superior femoral neck bone defect: multimodal biomechanical and clinical study.Frontiers in surgery · 2026
- TP4.4 A Three-Dimensional Morphological Analysis of Cam Lesions in Femoroacetabular Impingement Syndrome Yields Three Distinct Phenotypes: Moving Beyond the Alpha AngleJournal of Hip Preservation Surgery · 2026
- A primary study of inverted right-angle configuration cannulated screws fixation of femoral neck fracture based on elaborated 3-dimensional features of the femoral head and neck region.PloS one · 2026
- Use of patient-specific instrumentation for hip resurfacing arthroplasty: Measurement of accuracy using pre and postoperative CT-scan.Orthopaedics & traumatology, surgery & research : OTSR · 2026
- Preoperative measurement of femoral head using 3D-CT spherical fitting technique for the guidance of femoroplasty in arthroscopic treatment of cam-type femoroacetabular impingement: a cohort validation.Frontiers in medicine · 2026
- A retrospective cohort study comparing combined TiRobot navigation and percutaneous minimally invasive FNS fixation versus conventional open reduction FNS for femoral neck fractures.Journal of robotic surgery · 2026