Implementation of the first practical automatic micropulse blood pressure measurement device using the published revolutionary Newcastle technique.
Alan Murray, Dingchang Zheng, Clive J Griffiths, Chengyu Liu, David J Graham, Jeffrey Neasham, Adrian J Cossor
School of Engineering & Faculty of Medical Sciences, Newcastle University, Newcastle upon Tyne, UK. alan.murray@ncl.ac.uk. · Centre for Healthcare and Community Transformation, Coventry University, Coventry, UK.
Key points
From the paper's conclusion- We have now developed a first automated device using our technique and have compared its accuracy with current gold standard auscultatory measurements in 32 subjects.
Abstract
True accuracy for blood pressure measurement devices is required, but current techniques only estimate, deriving measurements from various private sets of clinical measurements. We previously reported a new technique that detects micropulse arterial opening, present only for beats between systole and diastole, enabling direct determination of systolic and diastolic blood pressures (SBP & DBP). We have now developed a first automated device using our technique and have compared its accuracy with current gold standard auscultatory measurements in 32 subjects. The differences between our novel device and simultaneous auscultatory measurements were for SBP -2.5 ± 3.1 mmHg, and for DBP -1.0 ± 1.2 mmHg (mean ± standard deviation SD). Results were well within the clinical international standard's limits of 5 ± 8 mmHg.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Toward patient-centered blood pressure monitoring: validation of a novel low-pressure and rapid oscillometric method according to the ISO 81060-2:2018/Amd1:2020/Amd2:2024 standard.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- A dataset of paired auscultatory and oscillometric blood pressure measurements from routine clinical practice.Scientific data · 2026
- Toward patient-centered blood pressure monitoring: first evaluation of a low-pressure and rapid oscillometric method for reducing cuff-related measurement burden.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Validation of the ARM-30F-XL arm blood pressure monitor with multiple cuff sizes according to the International Standard ISO 81060-2:2018+Amd 2:2024.Blood pressure monitoring · 2026
- Clinical validation of MESI mTABLET BP according to ISO 81060-2:2018 + A1 + A2.Blood pressure monitoring · 2026
- Accuracy of cuffless blood pressure devices compared with invasive arterial line monitoring in high-acuity settings: a random-effects meta-analysisEuropean Heart Journal · 2026