Fabrication of photoresist SU-8 pyrolyzed carbon electrodes and their integration in a microfluidic hematocrit detection system.
Hwi Yong Lee, Jessika A Rogers, Chito Kendrick, Sanaz Habibi, Adrienne R Minerick
Department of Chemical Engineering, Michigan Technological University, 1400 Townsend Drive, Houghton, MI, 49931, USA. · E&I Engineer at Oji Fiber Solutions, Ngongotaha Valley, Bay of Plenty, Rotorua, New Zealand.
Key points
From the paper's conclusion- The electrodes were integrated into a PDMS-based microfluidic device for hematocrit detection by measuring current responses of red blood cells suspended in phosphate-buffered saline at 100 V for 30 s.
Abstract
Carbon microelectrodes are attractive for microfluidic biosensing due to their wide potential window, low overpotentials, chemical stability, and compatibility with low-cost microfabrication. Pyrolysis of photopatterned SU-8 offers a scalable route to carbon electrode fabrication; however, conventional two-step pyrolysis often leads to adhesion issues, geometric distortion, and limited surface area in low-aspect-ratio planar designs. These constraints restrict performance and broader adoption in analytical microdevices. Hematocrit determination, a clinically essential diagnostic measurement, requires electrodes with high reproducibility and robust electrochemical response. A simplified fabrication strategy that improves electrode surface area and performance while maintaining low-aspect-ratio planar geometries remains unresolved.We developed a three-step pyrolysis protocol that enables fabrication of low-aspect-ratio planar SU-8-derived carbon electrodes with both micro- and millimeter-scale features while improving structural integrity and surface morphology. The electrodes were integrated into a PDMS-based microfluidic device for hematocrit detection by measuring current responses of red blood cells suspended in phosphate-buffered saline at 100 V for 30 s. A linear correlation was observed between current and red blood cell concentration. The device achieved 3.6% precision and 3.8% accuracy for hematocrit determination, comparable to previously reported platinum-based systems (2.8% precision, 2.6% accuracy). Surface characterization via field-emission scanning electron microscopy and atomic force microscopy revealed a 25% increase in surface area relative to platinum electrodes, resulting from 2.7 times greater roughness and 58 times greater thickness. These structural enhancements reduced charge-transfer and concentration overpotentials, improving electrochemical performance.This work introduces a reproducible three-step pyrolysis method that overcomes long-standing limitations in the fabrication of low-aspect-ratio pyrolytic carbon electrodes. By increasing electroactive surface area without complex coatings or noble metals, the approach enables high-performance, low-cost microelectrodes suitable for integrated biosensing. The demonstrated equivalence to platinum electrodes in hematocrit detection highlights its translational potential for scalable diagnostic microdevices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Universal Parafilm®-Mediated Thermal Bonding for Gold Leaf Electrode Patterning and Microfluidic Assembly on Diverse Substrates.Micromachines · 2026
- Embedded Printing-Derived Pyrolytic Carbon Electrodes with Graphene-like Features for Neural Stem Cell Differentiation and Dopamine SensingbioRxiv · 2026
- A thread-based microfluidic electrochemical platform for whole-blood protein sensing using endogenous uric acid as a redox mediator.Lab on a chip · 2026
- Poly(Neutral Red)-Silver Nanorods-Carbon Nanotubes Composite-Based Ratiometric Electrochemical Sensor for Rapid Detection of Histamine in Crayfish.Foods (Basel, Switzerland) · 2026
- Screen-printed carbon electrodes: processing, interface reconstruction, performance and sustainabilityJournal of Electrochemical Science and Engineering · 2026
- MEC-BioSense: Material-Encoded Capillary Biointerface for Autonomous Whole-Blood Biomarker Sensing.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2026