Reproducibility of T2 relaxation time mapping in a phantom and human knee articular cartilage.
Henri P P Leskinen, Nea Pulkkinen, Antti Aarnio, Amir Esrafilian, Teija Sainio, Eveliina Lammentausta, Ville Kantola, Victor Casula,
Department of Technical Physics, University of Eastern Finland, Yliopistonranta 8, 70210, Kuopio, Finland. · Department of Medical Physics, Turku University Hospital and University of Turku, Turku, Finland.
Key points
From the paper's conclusion- Low-to-moderate inter-scanner variation for the knee ROIs, CV = 2.2% to 22.8%, CVRMS,1.5 T = 9.3%, CVRMS,3 T = 11.0%, was observed. The phantom measurements showed low inter-scanner variation (CVRMS, 1.5 T = 4.7%, CVRMS, 3 T = 2.4%) on the typical cartilage T2 range.
Abstract
OBJECTIVE: To quantify the reproducibility of knee articular cartilage T2 across a wide range of MR systems. MATERIALS AND METHODS: The reproducibility of T2 relaxation time mapping was assessed by measuring the knee of a volunteer (n1.5 T = 22, n3T = 13) and a System Standard 130 phantom (n1.5 T = 19, n3T = 12) in 38 different clinical MRI scanners at 16 imaging sites. Root-mean-square coefficient of variation (CVRMS) was used to quantify the reproducibility of T2 at the selected knee cartilage regions of interest (ROI) and the phantom T2 array vials. RESULTS: Low-to-moderate inter-scanner variation for the knee ROIs, CV = 2.2% to 22.8%, CVRMS,1.5 T = 9.3%, CVRMS,3 T = 11.0%, was observed. The phantom measurements showed low inter-scanner variation (CVRMS, 1.5 T = 4.7%, CVRMS, 3 T = 2.4%) on the typical cartilage T2 range. DISCUSSION: Reproducibility of quantitative T2 can vary between healthcare facilities and scanners but also between the cartilage compartments. The variation is sufficiently small to consider T2 mapping reproducible. Phantom measurements support this result.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Musculoskeletal MRI Relaxometry Phantom for T1, T2, and T1ρ Measurements.Journal of magnetic resonance imaging : JMRI · 2026
- Textural Features from Quantitative MRI Using T2 Relaxation Times Correlate with Histological Cartilage Degeneration: An Ex-Vivo Study.Cartilage · 2026
- In vivo T1 and T2 relaxation times of the rodent brain at 5.0 T: cross-species quantification on a clinical MRI platform.Magma (New York, N.Y.) · 2026
- Reproducibility of 3.0 T single-frequency cardiac MR elastography for quantification of left ventricular myocardial stiffness.Radiology advances · 2026
- Assessment of intra- and inter-observer variability in T1 relaxation time measurement for rectal cancerActa Radiologica · 2026
- Systematic Review of the Reliability and Reproducibility of Magnetisation Transfer Imaging in Human Cervical Spinal Cord.Journal of clinical medicine · 2026