In vivo T1 and T2 relaxation times of the rodent brain at 5.0 T: cross-species quantification on a clinical MRI platform.
Xinxin Zhao, Xin Xu, Wen Wang, Jiahui Cheng, Shiteng Suo, Shaofeng Duan, Bo Li, Yan Zhou
Department of Radiology, Renji Hospital Affiliated to Shanghai Jiao Tong University School of Medicine, No. 160 Pujian Rd.,Pudong New District, Shanghai, 200127, China. · Collaborative Innovation Department, United Imaging Healthcare, No. 2258 Chengbei Rd.,Jiading District, Shanghai, 201807, China.
Key points
From the paper's conclusion- Phantom R1 and R2 demonstrated strong linear relationships with CuSO4 · 5H2O concentration (R2 = 0.9763 and 0.9583, respectively; both P < 0.001). Regional scan-rescan ICCs were 0.89-0.94 for T1 and 0.88-0.94 for T2, with corresponding CVs of 2.4%-6.9% and 2.6%-5.5%, respectively. Interobserver reproducibility was excellent. Both species showed the longest relaxation times in cerebrospinal fluid and the shortest in the corpus callosum. Mice had significantly longer T1 values than rats in all evaluated regions (all P < 0.001), whereas a significant interspecies difference in T2 was observed only in cerebrospinal fluid (P < 0.001).
Abstract
OBJECT: To assess the technical reproducibility of quantitative T1 and T2 mapping and characterize regional brain relaxation values in healthy rats and mice using a clinical 5.0-T MRI system with dedicated rodent coils. MATERIALS AND METHODS: Twenty Sprague-Dawley rats and 20 C57BL/6 mice, with equal numbers of females and males, underwent baseline and next-day MRI. T1 and T2 were measured using B1-corrected variable flip angle mapping and a CPMG-based multislice multiecho sequence, respectively. Six brain regions were measured independently by two blinded readers. Technical performance was assessed using repeatedly scanned single-compartment and nine-concentration CuSO4 · 5H2O phantoms. Scan-rescan and interobserver reproducibility were evaluated using intraclass correlation coefficients (ICCs) and coefficients of variation (CVs). RESULTS: Phantom R1 and R2 demonstrated strong linear relationships with CuSO4 · 5H2O concentration (R2 = 0.9763 and 0.9583, respectively; both P < 0.001). Regional scan-rescan ICCs were 0.89-0.94 for T1 and 0.88-0.94 for T2, with corresponding CVs of 2.4%-6.9% and 2.6%-5.5%, respectively. Interobserver reproducibility was excellent. Both species showed the longest relaxation times in cerebrospinal fluid and the shortest in the corpus callosum. Mice had significantly longer T1 values than rats in all evaluated regions (all P < 0.001), whereas a significant interspecies difference in T2 was observed only in cerebrospinal fluid (P < 0.001). DISCUSSION: Clinical 5.0-T MRI provides reproducible, platform- and protocol-specific regional T1 and T2 measurements in healthy adult rats and mice. These reference measurements support the application of clinical 5.0-T systems in preclinical quantitative and translational MRI research.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multidimensional T 1 - T 2 Relaxation Imaging In Vivo on a Portable 0.064 T MRI Scanner.bioRxiv : the preprint server for biology · 2026
- Validation of 1 T MRI in Cuprizone-Induced Demyelination in C57BL/6 Mice: Comparison with Histology and 7 T MRI.Molecular imaging and biology · 2026
- Resomapper: An Integrated Workflow for Quantitative Multiparametric MRI in Preclinical Studies.Journal of neuroscience methods · 2026
- Musculoskeletal MRI Relaxometry Phantom for T1, T2, and T1ρ Measurements.Journal of magnetic resonance imaging : JMRI · 2026
- Myocardial T1 and T2 Mapping at 5.0T: Feasibility and Comparison with 3.0-T MRI.Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance · 2026
- Reproducibility of T2 relaxation time mapping in a phantom and human knee articular cartilage.Magma (New York, N.Y.) · 2026