Journal of imaging informatics in medicine

Propagation of Low-Dose-Induced Noise Characteristics Across Deep Learning Layers in CT Imaging.

Hendrik Erenstein, Wim P Krijnen, Annemieke van der Heij-Meijer, Peter van Ooijen

Department of Medical Imaging and Radiation Therapy, Hanze University of Applied Sciences, Groningen, 9714 CA, The Netherlands. h.erenstein@pl.hanze.nl. · Research Group Healthy Ageing, Allied Health Care and Nursing, Hanze University of Applied Sciences, Groningen, 9714 CA, The Netherlands.

Computational

Key points

From the paper's conclusion
  • Feature analysis revealed progressive convergence of noise characteristics across deep learning layers, with batch normalization mitigating amplification observed after convolution and ReLU activation.

Abstract

Dose reduction in computed tomography (CT) imaging impacts diagnostic accuracy for human observers and potentially artificial intelligence (AI) systems. As noise propagates through an AI model, its characteristics change, possibly influencing model performance. This study aims to develop a configurable low-dose CT simulation tool in image-space and analyze noise behavior in deep learning (DL) layers. The noise characteristics were quantified using the noise power spectrum (NPS) within homogeneous phantom regions. NPS area under the curve (intensity) and median frequency (texture) were reported. Low-dose CT images were simulated using a custom program that was calibrated for configurable noise intensity and texture. Noise was generated at texture levels. Pre-trained U-Net features were extracted from reference and simulated low-dose images, enabling NPS-based analysis of noise propagation across deep learning layers. NPS from simulated low-dose CT images matched reference NPS for various intensity and texture values. Feature analysis revealed progressive convergence of noise characteristics across deep learning layers, with batch normalization mitigating amplification observed after convolution and ReLU activation. Differences in noise texture between low-dose simulations converge along the U-Net contracting path. The image-space simulation tool offers a viable alternative for researchers lacking access to raw data, supporting future AI-focused dose optimization studies. However, whether these outcomes translate to segmentation performance remains unclear and warrants further investigation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Propagation of Low-Dose-Induced Noise Characteristics Across Deep Learning Layers in CT Imaging — Science Skim