Association of Implantation Depth With Hemodynamics and Mechanical Consequences in Small Valve-In-Valve TAVR.
Bo Fu, Fenlong Xue, Yanhe Ma, Zhengda Chen, Shihong Liu, Hongyue Sun, Zhigang Guo, Qingliang Chen,
Department of Cardiovascular Surgery, Chest Hospital, Tianjin University, Tianjin, China. · Clinical School of Thoracic, Tianjin Medical University, Tianjin, China.
Key points
From the paper's conclusion- In small ViV-transcatheter aortic valve replacement, the hemodynamic effects of implantation depth are valve-combination specific. A depth-window strategy may improve acute hydrodynamic performance, but clinical translation requires balancing hemodynamic benefit against leaflet deformation and durability-related uncertainty.
Abstract
BACKGROUND: Valve-in-valve (ViV) transcatheter aortic valve replacement in small surgical bioprostheses is often limited by residual gradients and prosthesis-patient mismatch. Implantation depth is operator-controlled, but depth guidance for different valve combinations remains limited. OBJECTIVES: The authors aimed to evaluate the effects of implantation depth on ViV hemodynamics, leaflet deformation, and pinwheeling in small surgical bioprostheses and to assess the mechanical consequences of THV underexpansion using accelerated wear testing. METHODS: A 21-mm self-expanding transcatheter heart valve was implanted in five 21-mm surgical heart valves at multiple depths in a pulse duplicator. Mean gradient, effective orifice area, regurgitation volume, energy loss, leaflet deformation, and pinwheeling were assessed. Selected constrained ViV configurations underwent 200 million cycles of accelerated wear testing under ISO 5840 conditions. RESULTS: Implantation depth substantially influenced ViV performance. In 3 porcine surgical valves, the most favorable combined hemodynamic profile occurred within a moderate-to-deep depth window of approximately 13.0 to 14.5 mm, although the optimal value differed across individual hemodynamic and deformation endpoints. Excessive depth increased frame constraint, leaflet distortion, and pinwheeling. Two bovine surgical valves showed mixed or U-shaped responses, consistent with differences in true internal diameter and transcatheter valve oversizing. The selected ViV configurations completed 200 million accelerated wear cycles without gross structural failure under bench conditions. CONCLUSIONS: In small ViV-transcatheter aortic valve replacement, the hemodynamic effects of implantation depth are valve-combination specific. A depth-window strategy may improve acute hydrodynamic performance, but clinical translation requires balancing hemodynamic benefit against leaflet deformation and durability-related uncertainty.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Outcome of bioprosthetic valve fracture in small-valve ViV-TAVI proceduresEuropean Heart Journal · 2026
- Optimized Balloon-Expandable Transcatheter Aortic Valve Implantation and Clinical Outcomes.JACC. Advances · 2026
- Outcomes of Supra-Annular vs Intra-Annular Self-Expanding Valves in Valve-In-Valve Transcatheter Aortic Valve Replacement.JACC. Cardiovascular interventions · 2026
- Clinical Outcomes and Valve Performance of Intra-Annular Self-Expanding Transcatheter Aortic Valve in Valve-in-Valve Procedures: Insights From the NAViV Registry.Journal of the American Heart Association · 2026
- Impact of Aortic Annulus Size on Clinical and Hemodynamic Outcomes of Transcatheter Aortic Valve Implantation With the Navitor Valve.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Latest-iteration self-expanding vs balloon-expandable transcatheter heart valves for failed surgical bioprostheses: The INNOVATE registry.European heart journal. Valvular and structural heart disease · 2026