On Balancing Sharpness of Excitation Patterns, Loudness, and Potential Effects on Speech Perception with Cochlear-Implant Current-Focussing Strategies.
François Guérit, Robert P Carlyon
Cambridge Hearing Group, MRC Cognition and Brain Sciences Unit, University of Cambridge, 15 Chaucer Road, Cambridge, CB2 7EF, UK. Francois.Guerit@mrc-cbu.cam.ac.uk. · Cambridge Hearing Group, MRC Cognition and Brain Sciences Unit, University of Cambridge, 15 Chaucer Road, Cambridge, CB2 7EF, UK.
Key points
From the paper's conclusion- Cochlear implant (CI) users struggle to understand speech under noisy conditions. This has long been attributed to channel interactions: a given neuron not only responds to electrical stimulation from the closest electrode but also is sensitive to stimulation from electrodes located further away. Shaping the stimulation voltage into a very sharp profile (e.g. using tripolar instead of the more common monopolar stimulation) has been proposed as a solution, only with limited success. We argue this is due to several factors: the strong correlation between envelopes of adjacent CI channels with speech sounds, the necessity to recruit enough neurons for sufficient loudness, and, at loud levels, the similarity between monopolar and sharp stimulation output voltages at locations that are at least 3-4 electrodes away from a given stimulation electrode.
- We describe a "Hilltop" method to effectively reduce the voltage more than 3-4 electrodes away from a given electrode, while keeping the same percept of place pitch as with monopolar stimulation. We assess this method psychophysically in human CI users (N = 7 males, 9 females) and with computational modelling.
- Based on computational modelling and on psychophysical measures of charge summation, we show that the voltage difference between monopolar and Hilltop stimulation persists at the level of the spiral ganglion neurons.
Abstract
PURPOSE: Cochlear implant (CI) users struggle to understand speech under noisy conditions. This has long been attributed to channel interactions: a given neuron not only responds to electrical stimulation from the closest electrode but also is sensitive to stimulation from electrodes located further away. Shaping the stimulation voltage into a very sharp profile (e.g. using tripolar instead of the more common monopolar stimulation) has been proposed as a solution, only with limited success. We argue this is due to several factors: the strong correlation between envelopes of adjacent CI channels with speech sounds, the necessity to recruit enough neurons for sufficient loudness, and, at loud levels, the similarity between monopolar and sharp stimulation output voltages at locations that are at least 3-4 electrodes away from a given stimulation electrode. METHODS: We describe a "Hilltop" method to effectively reduce the voltage more than 3-4 electrodes away from a given electrode, while keeping the same percept of place pitch as with monopolar stimulation. We assess this method psychophysically in human CI users (N = 7 males, 9 females) and with computational modelling. RESULTS: Based on computational modelling and on psychophysical measures of charge summation, we show that the voltage difference between monopolar and Hilltop stimulation persists at the level of the spiral ganglion neurons. CONCLUSIONS: The results warrant investigations with more complex, multichannel stimuli. We also describe possible limitations of this strategy, including the presence of sidelobes of opposite effective polarity in some participants, and suggest ways to mitigate their effects.
Read the paper
Discussion · 1
Science Skim's AI key points, plus readers' public thoughts
- Cochlear implant (CI) users struggle to understand speech under noisy conditions. This has long been attributed to channel interactions: a given neuron not only responds to electrical stimulation from the closest electrode but also is sensitive to stimulation from electrodes located further away. Shaping the stimulation voltage into a very sharp profile (e.g. using tripolar instead of the more common monopolar stimulation) has been proposed as a solution, only with limited success. We argue this is due to several factors: the strong correlation between envelopes of adjacent CI channels with speech sounds, the necessity to recruit enough neurons for sufficient loudness, and, at loud levels, the similarity between monopolar and sharp stimulation output voltages at locations that are at least 3-4 electrodes away from a given stimulation electrode.
- We describe a "Hilltop" method to effectively reduce the voltage more than 3-4 electrodes away from a given electrode, while keeping the same percept of place pitch as with monopolar stimulation. We assess this method psychophysically in human CI users (N = 7 males, 9 females) and with computational modelling.
- Based on computational modelling and on psychophysical measures of charge summation, we show that the voltage difference between monopolar and Hilltop stimulation persists at the level of the spiral ganglion neurons.
Readers' thoughts
No reader thoughts yet.
Related papers
- Effects of pulse amplitude, pulse phase duration, and stimulation mode on loudness growth in cochlear implant users.Hearing research · 2026
- Improving Spectral Resolution via Simultaneous Extracochlear and Apical Ground Contacts: A Neural Simulation StudyarXiv · 2026
- Multichannel optical cochlear implants enable spectrally distinct auditory activity.EMBO molecular medicine · 2026
- Spectral sparsity enables discrimination of complex-tone triads in cochlear implant listeners.The Journal of the Acoustical Society of America · 2026
- Spectro-Temporal Modulation Sensitivity Prospectively Predicts Speech-in-Noise Recognition in Cochlear Implant Users.Ear and hearing · 2026
- CORSICA: Reproducible suppression of cochlear implant artifacts in EEG evoked by continuous speechJournal of Neural Engineering · 2026