Effects of pulse amplitude, pulse phase duration, and stimulation mode on loudness growth in cochlear implant users.
John J Galvin, David Bakhos, Matthew Robier, Bradford Backus, Emily Spitzer, David M Landsberger
House Institute Foundation, Los Angeles, CA, USA; CHRU de Tours, service ORL et Chirurgie Cervico-Faciale, 2 boulevard Tonnellé, 37044, Tours, France. Electronic address: Jgalvin@hifla.org. · House Institute Foundation, Los Angeles, CA, USA; CHRU de Tours, service ORL et Chirurgie Cervico-Faciale, 2 boulevard Tonnellé, 37044, Tours, France.
Key points
From the paper's conclusion- Loudness growth was steeper with focused than with MP stimulation, but only in OM users (p < 0.05), with no effect of stimulation mode in Cochlear users.
Abstract
In cochlear implants (CIs), loudness grows with increasing pulse amplitude (PA) or pulse phase duration (PPD). With a fixed PPD, the current field and the spread of excitation (SOE) broaden with increasing PA; with a fixed PA, the SOE increases with PPD. Current spread has been shown to be greater with monopolar (MP) than with focused stimulation. The effects of level coding and stimulation mode may reflect different aspects of recruitment. In the present study, we investigated the effects of stimulation mode (MP or focused), level coding (PA or PPD), and cochlear region on loudness growth in Cochlear and Oticon Medical (OM) device users. All charge levels were log-transformed before analysis. Loudness growth was steeper with PA than with PPD coding (p < 0.05). Loudness growth was steeper with focused than with MP stimulation, but only in OM users (p < 0.05), with no effect of stimulation mode in Cochlear users. Loudness growth was steeper in the basal than in the apical or middle regions of the cochlea (p < 0.05). The variability in loudness growth across electrodes was larger with PA than with PPD coding, but only in OM users (p < 0.05), and larger with focused than with MP stimulation, but only in Cochlear users (p < 0.05). PPD level coding may benefit CI users in terms of slower loudness growth (which may improve intensity resolution), reduced current (which may reduce electrode interaction), and less variable loudness growth across electrodes (which may better represent the acoustic input signal).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Relationships between ECAP amplitude growth functions and perceptual growth in cochlear implant users: Implications for individualized cochlear implant fitting.Hearing research · 2026
- Longitudinal Analysis of Cochlear Implant Stimulation Levels in the First Year: Optimizing Adult Fitting.Ear and hearing · 2026
- On Balancing Sharpness of Excitation Patterns, Loudness, and Potential Effects on Speech Perception with Cochlear-Implant Current-Focussing Strategies.Journal of the Association for Research in Otolaryngology : JARO · 2026
- Optimisation of and mechanisms behind neural distortion responses obtained with cochlear-implant stimulation by interleaved pulse trainsmedRxiv · 2026
- Pulse-rate effects on single- and multi-channel envelope processing and consonant identification in cochlear implant users.Hearing research · 2026
- Characterizing the perceptual effect of changing phase duration: Is it place-pitch?JASA express letters · 2026