Biochemical reaction-regulated electrocatalytic bipolar electrochemiluminescence platform with immobilization-free and exogenous electroactive-species-free sensing for cholesterol detection.
Mengyin Yang, Weiqiang Yang, Chengliang Ni, Zhiping Song, Xiaoping Chen, Qingxiang Wang, Jiancong Ni, Zhenyu Lin
Fujian Provincial Key Laboratory of Modern Analytical Science and Separation Technology, Fujian Provincial Key Laboratory of Pollution Monitoring and Control, College of Chemistry, Chemical Engineering and Environment, Minnan Normal University, 36 Xianqian Street, 363000, Zhangzhou, Fujian, P. R. China. · Zhangzhou Municipal Hospital of Fujian Province and Zhangzhou Affiliated Hospital of Fujian Medical University, Zhangzhou, 363000, China.
Key points
From the paper's conclusion- In addition, no externally added electroactive species are required in the sensing cell, while the Ru(bpy)32+/TPrA ECL system is confined to the reporting cell.
Abstract
A biochemical reaction-regulated electrocatalysis strategy for quantitative cholesterol detection using platinum nanoparticle-functionalized indium tin oxide electrodes for electrochemiluminescence. In this system, cholesterol is oxidized by cholesterol oxidase to generate H2O2 in situ, which serves as an endogenous electroactive species and undergoes reduction at the BPE cathode. According to the charge balance principle, the cathodic reaction drives the anodic oxidation of the Ru(bpy)32+/TPrA system, generating an ECL signal that correlates quantitatively with cholesterol concentration. PtNPs at the BPE cathode enhance the electrocatalytic reduction of H2O2 and accelerate interfacial electron transfer, leading to an amplified anodic ECL response. Under optimized conditions, the platform exhibits a linear range of 0.1-10 mM with an extrapolated detection limit of 98.4 µM. The proposed sensing strategy was successfully applied to cholesterol detection in human serum samples, demonstrating good accuracy and anti-interference capability. Notably, the ChOx-catalyzed oxidation of cholesterol proceeds homogeneously in the sensing solution, while no enzyme-immobilization step is required during sensor preparation. In addition, no externally added electroactive species are required in the sensing cell, while the Ru(bpy)32+/TPrA ECL system is confined to the reporting cell. This sensing/reporting separation effectively minimizes cross-interference.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Highly efficient electrochemical detection of cholesterol via hydrophobic matching host-guest recognition on bimetallic nanoparticle support.Talanta · 2026
- Fabrication of Pt-decorated MXene-chitosan nanocomposite electrodes for non-enzymatic electrochemical detection of biomolecules.RSC advances · 2026
- Electrochemiluminescence Biosensor Based on Dual-Signal Regulation of Gold Nanoclusters for Amyloid-β Protein DetectionAnalytical Chemistry · 2026
- Multifunctional PTCA/GO for a Glucose-Driven Electrochemiluminescence BiosensorNano Letters · 2025
- Synergistic catalysis of MoC nanoclusters/Co single-atom sites for H₂O₂-free electrochemiluminescence detection of ascorbic acid.Mikrochimica acta · 2026
- Electrochemical cardiac troponin I immunosensor platform integrated with electrochemically deposited Au nanoparticles decorated PtNi bimetallic sponge-like structures.Bioelectrochemistry (Amsterdam, Netherlands) · 2026