Adverse events and device failures of the transoral incisionless fundoplication system: a MAUDE database analysis.
Gedion Yilma Amdetsion, Tewodros Ayele, Roberto Oleas, Fnu Vikash, Hiwot Tebeje, Ayokunle T Abegunde
Internal Medicine, Department of Medicine, John H. Stroger Hospital of Cook County, 1969 W. Ogden Avenue, Chicago, IL, 60612, USA. Gedion.amdetsion@gmail.com. · Division of Gastroenterology and Hepatology, University of Iowa Hospitals and Clinics, Iowa, IA, USA.
Key points
From the paper's conclusion- MAUDE reports have risen sharply since 2020, dominated by severe patient harm, particularly perforation, pulmonary complications, bleeding, and mechanical failures. Despite MAUDE's limitations, the predominance of severe events warrants procedural standardization, credentialing pathways, and prospective safety registries as TIF/cTIF use expands.
Abstract
BACKGROUND: Transoral incisionless fundoplication (TIF) using the EsophyX device, with or without concomitant laparoscopic hiatal hernia repair (cTIF), is increasingly used for GERD. Postmarketing safety data are limited. We analyzed the FDA MAUDE database to characterize adverse events and device failures. METHODS: We extracted MDRs for the EsophyX device family and SerosaFuse fastener kit submitted through March 2026. Duplicates were removed by MDR identifier and narrative matching. Reports were classified for patient adverse events and device failures; severity was graded using a modified Cotton lexicon. RESULTS: After deduplication, 192 unique MDRs were analyzed (130 Injury, 60 Malfunction, 2 Death). Annual reports rose from 8.8/year in 2016-2019 to 26, 40, and 33 in 2023-2025. Adverse events occurred in 138 reports (71.9%) and device failures in 88 (45.8%). Most frequent events were perforation (58.0%), esophageal/mucosal injury (27.5%), pulmonary/pleural complications (21.0%), pain (21.0%), infection/sepsis (20.3%), and bleeding (19.6%). Five procedure-related deaths (3.6% of patient-event reports) were identified, of which only two were filed by FDA under the Death classification. By Cotton severity, 49.3% were severe and 3.6% fatal. Surgical intervention occurred in 34.4%, hospitalization in 29.2%, endoscopic rescue in 16.7%. Perforation was reported in cTIF in 56.6% and 30.3% TIF alone. Dominant failures were insertion/advancement/removal difficulty (30.7%), component detachment (23.9%), breakage (23.9%), and mechanical jam (22.7%). CONCLUSIONS: MAUDE reports have risen sharply since 2020, dominated by severe patient harm, particularly perforation, pulmonary complications, bleeding, and mechanical failures. Despite MAUDE's limitations, the predominance of severe events warrants procedural standardization, credentialing pathways, and prospective safety registries as TIF/cTIF use expands.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Early Post-Market Surveillance of Bimodal Neuromodulation for Tinnitus.Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery · 2026
- Safety Profile and Adverse Event Management in Flexible Endoscopic Treatment of Zenker's Diverticulum: A Systematic Review.Medicina (Kaunas, Lithuania) · 2026
- Adverse event reporting patterns for endodontic files in the FDA MAUDE database: a postmarket surveillance analysis, 2015-2025.Journal of endodontics · 2026
- Post-market individual-report phenotypes after pulsed-field ablation for atrial fibrillation in a food and drug administration database.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- All-Inside Meniscus Repair Systems Failure Analysis Shows Device-Related Adverse Events Differ by Device Design Type and Show Varied Reporting Delays by Manufacturer.Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association · 2026
- Reported device problems and patient outcomes in da Vinci robotic robotic gastric and upper-gastrointestinal procedures: an FDA MAUDE database analysis.Journal of robotic surgery · 2026