Comparison of outcomes after transcatheter patent ductus arteriosus closure in infants weighing ≤ 1.5 kg and > 1.5-4 kg.

Katia Librandi, Carles Bautista-Rodriguez, Gianfranco Butera, Giuseppe Annoni, Paul Padovani, Sandrine Foldvari, Alban-Elouen Baruteau, Alain Fraisse

Pediatric Cardiology, Royal Brompton & Harefield NHS Foundation Trust, Sydney Street, London, SW3 6NP, UK. · Cardiology, Cardiac Surgery and Heart Lung Transplantation, Bambino Gesù Hospital and Research Institute, IRCCS, Rome, Italy.

Observational · n=358

Key points

From the paper's conclusion
  • Transcatheter PDA closure is safe and effective in infants ≤ 4 kg, but the > 1.5-4 kg group has a higher risk of major adverse events, particularly device embolization. Better patient selection and dedicated devices may improve safety.

Abstract

UNLABELLED: Transcatheter closure of patent ductus arteriosus (PDA) with the Amplatzer Piccolo Occluder is safe and effective in infants ≤ 2 kg. In larger infants, off-label devices are often required, but outcome data are limited. We evaluated outcomes in infants ≤ 4 kg, comparing the ≤ 1.5 and > 1.5-4 kg groups. This retrospective, international, multicenter study included 358 infants who underwent transcatheter PDA closure at 5 tertiary centers between 2016 and 2024: 191 weighed ≤ 1.5 kg (group 1) and 167 > 1.5-4 kg (group 2). Median age was 39 days (interquartile range 25-61) and median weight was 1.48 (1.1-2.2) kg; 93.6% were preterm. Procedural success was higher in group 1 (98.4% vs 94.0%; p = 0.044). Major adverse events were more frequent in group 2 (9.6% vs 3.1%; p = 0.0144; odds ratio = 3.27; 95% confidence interval, 1.25-8.55). Device embolization occurred only in group 2 (5.4% vs 0%; p < 0.001; odds ratio = 22.96; 95% confidence interval, 1.33-397.51). The Amplatzer Piccolo Occluder was used in 97.3% of group 1 vs 69.5% of group 2. CONCLUSION: Transcatheter PDA closure is safe and effective in infants ≤ 4 kg, but the > 1.5-4 kg group has a higher risk of major adverse events, particularly device embolization. Better patient selection and dedicated devices may improve safety. WHAT IS KNOWN: • Transcatheter PDA closure with the Amplatzer Piccolo Occluder is safe and effective in very small infants. • In larger infants, alternative devices are often used off-label, but evidence on procedural outcomes remains limited. WHAT IS NEW: • Infants weighing >1.5-4 kg had lower procedural success and more major adverse events than those weighing ≤1.5 kg. • Device embolization occurred exclusively in the >1.5-4 kg group, supporting careful anatomical assessment and the development of dedicated devices.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers