Validation of a neonatal computational anthropomorphic phantom using Monte Carlo simulation.
Beatriz Hellen A Pereira, Franklin S S Leite, Alberico B Carvalho Junior, Lucio P Neves, Ana Paula Perini, Arthur S B Z Alves, William S Santos
Postgraduate Program in Biomedical Engineering (PPGEB), Faculty of Electrical Engineering, Federal University of Uberlândia, 2121 João Naves de Ávila Avenue, Santa Mônica, Uberlândia, MG 38408-100, Brazil. · Department of Physics, Federal University of Sergipe, Marcelo Deda Chagas Avenue, Rosa Elze, São Cristóvão, SE 49107-230, Brazil.
Key points
From the paper's conclusion- This study presents the validation of a neonatal computational anthropomorphic phantom developed by the Federal University of Sergipe (UFS) using Monte Carlo simulations performed with the MCNPX 2.7.0 code.
Abstract
This study presents the validation of a neonatal computational anthropomorphic phantom developed by the Federal University of Sergipe (UFS) using Monte Carlo simulations performed with the MCNPX 2.7.0 code. The model was compared with the ICRP 143 reference phantom in terms of absorbed and effective dose for 10 irradiation geometries (0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°, ISO and ROT) and photon energies from 50 keV to 10 MeV. The results presented relative differences <5% for anteroposterior irradiation, while small deviations were generally observed for isotropic (ISO) and rotational (ROT) geometries. Larger differences were observed at low energies and in directional geometries, reaching maximum values of ~30%-35%, particularly for posteroanterior irradiation, while minimum differences were <1% at higher energies. The same was observed for effective dose results. Overall, the differences decreased with increasing photon energy, confirming the reliability of the UFS phantom for neonatal dosimetry and radiation protection applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dosimetric impact of computational phantom selection in passive scattering proton therapy: Development and validation of an extended MIRD prostate phantom using Geant4 Monte Carlo simulation.Radiation and environmental biophysics · 2026
- Fetal specific absorbed fractions of ICRP pregnant-female mesh phantoms: methodology and dosimetric trends.Physics in medicine and biology · 2026
- Impact of phantom design on fetal radiation exposure assessment in pregnant women undergoing head and neck 6MV photon radiotherapy.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Electron specific absorbed fractions for the ICRP family of mesh-type reference computational phantoms.Physics in medicine and biology · 2026
- Photon and electron specific absorbed fractions to pregnant female organs from maternal and fetal sources.Radiation and environmental biophysics · 2026
- Development and validation of a pregnant phantom with adjustable fetal presentation for region-specific fetal radiation dose assessment during pelvic radiography.Journal of applied clinical medical physics · 2026