A feasible and cost-effective method to induce distal radius fractures in full-length forearm human cadaveric specimens.
Julius Krüger, Mads Terndrup, Morten Grove Thomsen, Arvind von Keudell, Jonas Askø Andersen, Tazio Maleitzke
Trauma Orthopaedic Research Copenhagen Hvidovre (TORCH), Department of Orthopaedic Surgery, Copenhagen University Hospital - Amager and Hvidovre, Hvidovre, Denmark. · Exo360, Copenhagen, Denmark.
Key points
From the paper's conclusion- This study presents a feasible, cost-effective method to induce DRFs in tissue-intact, full-length human cadaveric forearm specimens. The setup consistently produces DRFs, predominantly AO/OTA type C, which can, with certain limitations, be utilized for biomechanical testing of immobilization devices, while also being uited for surgical and anatomical education.
Abstract
INTRODUCTION: Reported rates of loss of reduction in nonoperatively treated distal radius fractures (DRF) range from 17 to 71%. Thus, realistic biomechanical models are warranted to test novel immobilization devices that prevent loss of reduction in nonoperative fracture management. Currently available human cadaveric specimen models have several limitations: (i) disrupted soft tissue integrity, (ii) difficulty in replicating complex fracture patterns, and (iii) high technical demands and costs. Therefore, the aim of this study was to establish a feasible and accessible method for inducing DRFs, that accurately reflects in-vivo biomechanical fracture conditions in tissue-intact cadaveric specimens, suitable for testing of novel immobilization devices. MATERIALS AND METHODS: Fifteen upper extremity cadaveric specimens were included. Specimens were prepared as full‑length forearm models. Using a repurposed strength training tower as a vertical drop tower, fractures were induced by proximally administering 120-135 J impacts. CT scans were performed pre- and post-fracture induction. All fractures were classified according to the AO/OTA classification. RESULTS: DRFs were successfully induced in all 15 specimens. A single 120 J impact produced a DRF in 11 out of 15 cases (73%). A second impact was administered in 4 out of 15 specimens (27%). Eleven of the 15 induced DRFs (73%) were classified as AO/OTA type C fractures, most commonly 2R3C3.2. Concomitant distal ulnar fractures were present in 6 cases, concomitant carpal fractures in 9 specimens. The overall cost of the setup amounted to 947 €. CONCLUSIONS: This study presents a feasible, cost-effective method to induce DRFs in tissue-intact, full-length human cadaveric forearm specimens. The setup consistently produces DRFs, predominantly AO/OTA type C, which can, with certain limitations, be utilized for biomechanical testing of immobilization devices, while also being uited for surgical and anatomical education.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biomechanical comparison of palmar plate plus headless compression screw versus radiopalmar double plating in AO/OTA 23-C2.1 distal radius fracturesArchives of Orthopaedic and Trauma Surgery · 2026
- Distal radius fractures reduction and immobilization using a novel device in older adults in emergency department.Frontiers in surgery · 2026
- Radial head arthroplasty restores axial construct stiffness in a Monteggia-like injury model: a biomechanical study.Archives of orthopaedic and trauma surgery · 2026
- Internal Fixation Bone Mass Index (IFBM) and Biomechanical Performance of Different Fixation Methods for Herscovici Type B and Type C Medial Malleolar Fractures.The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons · 2026
- A novel method of manipulation and immobilization with the wrist dorsiflexed for distal radius fractures: short-term outcomes.Journal of orthopaedic surgery and research · 2026
- Comparative biomechanical study of a novel variable diameter circular external fixation device (VDCF) for stabilizing a simulated tibial fracture gap in bovine cadaver specimens.Veterinary surgery : VS · 2026