The effect of magnetic interference by magnetic resonance imaging on programmable shunt valves in patients with normal pressure hydrocephalus: a systematic review and pooled analysis.
Virensinh Rathod, Darius Kalasauskas, Florian Ringel, Harroop Bola, Daniele S C Ramsay, Karanjot Chhatwal, Srikar R Namireddy, Sree Kanakala,
Imperial Brain & Spine Initiative, Imperial College London, London, UK. · Department of Neurosurgery, LMU University Hospital, LMU Munich, Munich, Germany.
Key points
From the paper's conclusion- The pooled analysis included 475 NPH patients across 11 studies. Valve setting alteration was required in 44.2% of cases following 1.5T exposure and 55.2% after 3T exposure, with no statistically significant difference between field strengths (p > 0.05). Valve type analysis revealed that the Codman Hakim valve exhibited the highest incidence of setting changes (37.05% at 3T). However, differences in valve setting changes between manufacturers were not statistically significant (p > 0.05).
Abstract
INTRODUCTION: Normal pressure hydrocephalus (NPH) is a cause of neurological impairment, with programmable shunt valves serving as the standard treatment. However, magnetic interference can result in altered pressure settings, affecting valve functionality. RESEARCH QUESTION: This systematic review investigates the effects of magnetic resonance imaging (MRI) on programmable shunt valves in NPH patients. MATERIAL AND METHODS: Conducted in adherence to PRISMA guidelines, an extensive literature search across PubMed and Embase databases from January 2000 to January 2025, was performed. Studies evaluating the impact of MRI at 1.5T and 3T field strengths on valve settings were included. Valve reprogrammability and outcomes were analysed, with statistical comparisons across MRI strengths and valve types using Mann-Whitney U and ANOVA tests. RESULTS: The pooled analysis included 475 NPH patients across 11 studies. Valve setting alteration was required in 44.2% of cases following 1.5T exposure and 55.2% after 3T exposure, with no statistically significant difference between field strengths (p > 0.05). Valve type analysis revealed that the Codman Hakim valve exhibited the highest incidence of setting changes (37.05% at 3T). However, differences in valve setting changes between manufacturers were not statistically significant (p > 0.05). DISCUSSION AND CONCLUSION: Magnetic interference from MRI can disrupt programmable shunt valves in NPH patients. While 3T MRI shows a slightly higher incidence of valve setting changes than 1.5T, the difference is not significant. Further research is required to assess long- term clinical implications and standardize MRI protocols to mitigate variability in outcomes. CLINICAL TRIAL NUMBER: Not applicable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Programmable shunt valve stability after MRI: a 2-year institutional experience across valve types and field strengths.Child's nervous system : ChNS : official journal of the International Society for Pediatric Neurosurgery · 2026
- Magnetic interference and programmable valves: navigating hidden risks in everyday life and clinical settings.Child's nervous system : ChNS : official journal of the International Society for Pediatric Neurosurgery · 2026
- Impact of Valve Selection on Clinical Outcomes in Idiopathic Normal Pressure Hydrocephalus: A Retrospective Cohort Study.World neurosurgery · 2026
- Optimization of phase-contrast MRI shunt flow measurement across multiple scanners for the management of hydrocephalus.Journal of neurosurgery. Pediatrics · 2026
- Transcranial doppler ultrasonography in normal pressure hydrocephalus: a systematic review and exploratory meta-analysis of diagnostic, prognostic, and monitoring roles.Neurosurgical review · 2026
- Motor network neurophysiological (TMS-EEG) markers as predictors of outcomes of ventriculoperitoneal shunt sustained for over a year in normal pressure hydrocephalus.Fluids and barriers of the CNS · 2026