Intraarticular calcaneal fractures: biomechanical analysis of two fixation constructs.
Franziska Ziegenhain, Ivan Zderic, Robert Zondervan, R Geoff Richards, Boyko Gueorguiev, Stefan Rammelt, Tim Schepers, Michael Swords
AO Research Institute Davos, Davos, Switzerland. · AO Research Institute Davos, Davos, Switzerland. ivan.zderic@aofoundation.org.
Key points
From the paper's conclusion- Biomechanically, fixation implementing a small low-profile 2.7 mm variable-angle locking anterolateral calcaneal plate supplemented with two additional independent cortex screws demonstrated stability comparable to, and in certain respects superior to that achieved with a small 3.5 mm locking calcaneal plate in a standardized Sanders IIc fracture model. However, the findings of this laboratory study do not establish an isolated benefit of the 2.7 mm plate itself, nor do they demonstrate clinical superiority or support alterations to postoperative weight-bearing protocols.
Abstract
PURPOSE: The aim of this biomechanical study was to investigate the fixation stability of a small low-profile 2.7 mm variable-angle locking anterolateral calcaneal plate supplemented with two additional independent cortex screws versus a traditional small 3.5 mm locking calcaneal plate in a standardized Sanders IIc fracture model. METHODS: Sixteen fracture models cocomposed of three polymer blocks (fragments) each and simulating Sanders IIc intra‑articular calcaneal fractures were assigned to two groups. Group 1 (control) received a small 3.5 mm locking calcaneal plate, while Group 2 received a small 2.7 mm variable‑angle locking anterolateral calcaneal plate with two additional independent cortex screws. All constructs were instrumented using prefabricated pilot holes and standardized screw configurations. Biomechanical testing was conducted implementing a 3‑point-bending setup and progressively increasing cyclic loading to failure. Angular interfragmentary movements were quantified via motion tracking along with cycles to clinically relevant failure. RESULTS: Initial stiffness (N/mm) in Group 1 (65.2 ± 34.7) was significantly lower versus Group 2 (232.5 ± 48.6), p = 0.020. Although angular interfragmentary movements around the anteroposterior axis were significantly greater in Group 1 versus Group 2 (p < 0.001), Group 1 demonstrated significantly smaller angular interfragmentary movements around the mediolateral axis for the central-to-posterior and anterior-to-posterior fragment pairs (p ≤ 0.004), while no significant difference between the groups was identified with regard to the central-to-anterior fragment pair (p ≥ 0.377). Cycles and loads (N) to clinically relevant failure in Group 1 (3635 ± 176 and 92.7 ± 3.5) were significantly lower versus Group 2 (17433 ± 628 and 368.7 ± 12.6), p < 0.001. Failure mode in Group 1 involved plate deformation, whereas Group 2 failed through breakage or bending of the independent cortex screws. CONCLUSIONS: Biomechanically, fixation implementing a small low-profile 2.7 mm variable-angle locking anterolateral calcaneal plate supplemented with two additional independent cortex screws demonstrated stability comparable to, and in certain respects superior to that achieved with a small 3.5 mm locking calcaneal plate in a standardized Sanders IIc fracture model. However, the findings of this laboratory study do not establish an isolated benefit of the 2.7 mm plate itself, nor do they demonstrate clinical superiority or support alterations to postoperative weight-bearing protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Combo or perimeter plate? fixation strategy for intra-articular calcaneal fracture via the sinus tarsi approach: a retrospective cohort studyBMC Surgery · 2026
- Biomechanical comparison of palmar plate plus headless compression screw versus radiopalmar double plating in AO/OTA 23-C2.1 distal radius fracturesArchives of Orthopaedic and Trauma Surgery · 2026
- Closed percutaneous reduction versus mini-open sinus tarsi approach for fixation of displaced intra-articular calcaneal fractures: a systematic review and meta-analysis.European journal of orthopaedic surgery & traumatology : orthopedie traumatologie · 2026
- Biomechanical comparison of parallel versus convergent screw fixation in Zadek osteotomy.European journal of orthopaedic surgery & traumatology : orthopedie traumatologie · 2026
- Clinical application of robot-assisted fixation for calcaneal fractures: a meta-analysis with subgroup comparisons.Journal of robotic surgery · 2026
- Comparative finite element analysis of a novel distal tibiofibular locking plate versus conventional fixation methods for Tillaux-Chaput fractures.Frontiers in bioengineering and biotechnology · 2026