Liver T1, T2, T2*, and PDFF Mapping at 0.55 T Using Rosette MRF With Optimized Sequence Design and Deep Image Reconstruction.

Tom Griesler, Jesse Hamilton, Evan Cummings, Jacob Richardson, Andrew Whiteman, Maximilian Gram, Gastao Cruz, Hero Hussain,

Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA. · Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA.

Key points

From the paper's conclusion
  • CRLB-guided acquisition and multi-echo DIP reconstruction enable accurate and precise liver MRF T1/T2/T2*/PDFF mapping at 0.55 T in a single breath-hold.

Abstract

PURPOSE: Quantitative imaging on mid-field (0.55 T) MRI systems can suffer from poor precision due to low SNR. This work presents an MRF pulse sequence optimized with Cramer-Rao lower bound (CRLB) and combined with a multi-echo deep image prior (DIP) reconstruction for co-registered T1/T2/T2*/PDFF mapping in the liver at 0.55 T within a single breath-hold. METHODS: A 20 s 2D multi-echo rosette MRF breath-hold acquisition was implemented on a 0.55 T wide-bore system. The sequence design was optimized via pseudorandom search using a CRLB-based objective for T1/T2 encoding and compared with reference designs. Data were reconstructed using (i) an iterative low-rank (LR) reconstruction and (ii) a multi-echo DIP jointly modeling TR- and echo-domain dynamics including B0 effects. Water/fat separation preceded dictionary matching for T1/T2; curve fitting was used to calculate T2* and PDFF was derived from M0 with noise-bias correction. The method was evaluated in numerical simulations, NIST and fat-fraction phantom scans, and in vivo acquisitions in 10 healthy subjects. A linear mixed-effects model was used to evaluate the effects of sequence design and reconstruction method on mapping accuracy and precision. RESULTS: Optimized sequences achieved lower CRLB cost than reference designs and, in simulations, CRLB cost was correlated with T1/T2 ROI variability (R2 = 0.76 LR; R2 = 0.83 DIP). Phantom experiments showed agreement with reference values. In vivo, both sequence optimization and DIP reconstruction significantly reduced the standard deviation of all four parameters in liver ROIs. CONCLUSION: CRLB-guided acquisition and multi-echo DIP reconstruction enable accurate and precise liver MRF T1/T2/T2*/PDFF mapping at 0.55 T in a single breath-hold.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Liver T1, T2, T2*, and PDFF Mapping at 0.55 T Using Rosette MRF With Optimized Sequence Design and Deep Image Reconstruction — Science Skim