Liver T1, T2, T2*, and PDFF Mapping at 0.55 T Using Rosette MRF With Optimized Sequence Design and Deep Image Reconstruction.
Tom Griesler, Jesse Hamilton, Evan Cummings, Jacob Richardson, Andrew Whiteman, Maximilian Gram, Gastao Cruz, Hero Hussain,
Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA. · Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA.
Key points
From the paper's conclusion- CRLB-guided acquisition and multi-echo DIP reconstruction enable accurate and precise liver MRF T1/T2/T2*/PDFF mapping at 0.55 T in a single breath-hold.
Abstract
PURPOSE: Quantitative imaging on mid-field (0.55 T) MRI systems can suffer from poor precision due to low SNR. This work presents an MRF pulse sequence optimized with Cramer-Rao lower bound (CRLB) and combined with a multi-echo deep image prior (DIP) reconstruction for co-registered T1/T2/T2*/PDFF mapping in the liver at 0.55 T within a single breath-hold. METHODS: A 20 s 2D multi-echo rosette MRF breath-hold acquisition was implemented on a 0.55 T wide-bore system. The sequence design was optimized via pseudorandom search using a CRLB-based objective for T1/T2 encoding and compared with reference designs. Data were reconstructed using (i) an iterative low-rank (LR) reconstruction and (ii) a multi-echo DIP jointly modeling TR- and echo-domain dynamics including B0 effects. Water/fat separation preceded dictionary matching for T1/T2; curve fitting was used to calculate T2* and PDFF was derived from M0 with noise-bias correction. The method was evaluated in numerical simulations, NIST and fat-fraction phantom scans, and in vivo acquisitions in 10 healthy subjects. A linear mixed-effects model was used to evaluate the effects of sequence design and reconstruction method on mapping accuracy and precision. RESULTS: Optimized sequences achieved lower CRLB cost than reference designs and, in simulations, CRLB cost was correlated with T1/T2 ROI variability (R2 = 0.76 LR; R2 = 0.83 DIP). Phantom experiments showed agreement with reference values. In vivo, both sequence optimization and DIP reconstruction significantly reduced the standard deviation of all four parameters in liver ROIs. CONCLUSION: CRLB-guided acquisition and multi-echo DIP reconstruction enable accurate and precise liver MRF T1/T2/T2*/PDFF mapping at 0.55 T in a single breath-hold.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Water-T1, PDFF, and T2* mapping of the liver using chemical shift-encoded MOLLI.Abdominal radiology (New York) · 2026
- Optimal RF Phase Modulation for Improved Magnetic Resonance Fingerprinting T2 Encoding and Efficiency.Magnetic resonance in medicine · 2026
- Simultaneous T1, T2, R2*, PDFF, and triglyceride quantification in the head and neck using magnetic resonance fingerprinting on a 1.5T MR-Linac.Medical physics · 2026
- Free-Breathing PDFF and R2* Quantification at 0.55T: Reduced Respiratory Motion Sensitivity and a Time-Resolved 4D MRI ApproacharXiv · 2026
- Mapping R1 of fat, R1 of water, joint R2⁎, and proton density fat fraction at 3 T using the multi-echo MP2RAGE sequence.Magnetic resonance imaging · 2026
- Cross‐Vendor Validation of Proton Density Fat Fraction and T₁ Mapping Using a Combined Proton Density Fat Fraction— T1 PhantomJournal of Magnetic Resonance Imaging · 2026