Key points

From the paper's conclusion
  • Simulation, phantom, and in vivo studies demonstrate that optimized RF phase modulation improves MRF T2 encoding compared to unoptimized and FA-only optimized sequences, enabling shorter exams and greater T2 precision in the brain and pelvis at 3T and 0.55T.

Abstract

PURPOSE: To improve magnetic resonance fingerprinting (MRF) T2 encoding by combining phase-based T2 encoding with optimal experiment design. METHODS: Six hundred-timepoint MRF sequences with jointly optimized flip angle (FA) and quadratic RF phase increment schedules were generated by minimizing their Cramér-Rao Lower Bound for tissue dependent sets of T1 and T2. Sequences were optimized for 3T and 0.55T brain MRF and 0.55T prostate MRF and compared against FA-only optimized and unoptimized (600- and 1000-timepoint) sequences. Accuracy and precision were first evaluated using Monte-Carlo Bloch simulations in a digital phantom. Additional simulations identified a spiral interleaf ordering which minimized aliasing artifacts under phase modulation. A T1/T2 phantom was then used to validate simulation results. In vivo performance was assessed in nine healthy subjects, where precision was measured using bootstrap Monte-Carlo experiments with 50 pseudo-replicas. RESULTS: Jointly optimized sequences were found to produce T2 maps with significantly greater accuracy and precision than unoptimized and FA-only optimized sequences in simulation and phantom experiments (p < 0.001). A Step 5 (37.5° between timepoints) spiral interleaf ordering minimized aliasing under phase modulation, while maintaining comparable performance without. In vivo testing in the brain and prostate revealed higher T2 values, reduced T2 coefficients of variation, and lower B1 + sensitivity than unoptimized and FA-only optimized sequence variants. CONCLUSION: Simulation, phantom, and in vivo studies demonstrate that optimized RF phase modulation improves MRF T2 encoding compared to unoptimized and FA-only optimized sequences, enabling shorter exams and greater T2 precision in the brain and pelvis at 3T and 0.55T.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Optimal RF Phase Modulation for Improved Magnetic Resonance Fingerprinting T2 Encoding and Efficiency — Science Skim