From Breath-Hold to Free-Breathing: Possibilities in High-Resolution Quantitative Susceptibility Mapping (QSM) of the Kidney With Respiratory Gating.
Matthias Deitermann, Birte Sehlmeyer, Thomas Andreas Thiel, Hans-Jörg Wittsack, Alexandra Ljimani, Anja Müller-Lutz, Gerald Antoch, Eric Bechler
Department of Diagnostic and Interventional Radiology, Medical Faculty and University Hospital Düsseldorf, Heinrich-Heine-University Düsseldorf, Germany.
Key points
From the paper's conclusion- Free-breathing QSM protocols in the human kidney using respiratory gating show equal susceptibility results compared to breath-hold acquisition while increasing clinical feasibility. A free-breathing protocol allows for the measurement of a larger subject pool and enables an increase in critical sequence parameters, such as echo-time and resolution. This could improve the accuracy of susceptibility calculations in the human abdomen.
Abstract
PURPOSE: This study aimed to assess the feasibility of QSM in the human kidney during free-breathing using respiratory gating for motion compensation. METHODS: Renal imaging of 15 healthy volunteers was performed using an axial multi-echo GRE sequence on a 3 T MRI scanner. Each subject was scanned once during breath-hold and once during free-breathing with respiratory gating based on a respiratory cushion. Five subjects were repositioned and scanned again to test for reproducibility. Additionally, one subject was measured in axial and coronal direction maximizing the feasible resolution during free-breathing. All phase data was unwrapped, the background field removed, and the inverse problem solved to create susceptibility maps. For statistical analysis, the mean susceptibility values of one kidney were analyzed across three consecutive slices and referenced to the susceptibility of the paravertebral muscle. RESULTS: The investigation revealed no statistically significant differences between the breath-hold and respiratory gated sequence. The mean susceptibility values recorded in the right kidney were 0.03 ± 0.06 ppm for the breath-hold and 0.03 ± 0.07 ppm for the free-breathing. Bland-Altman analysis demonstrated good reproducibility for both methods. The high-resolution measurements enabled by gated breathing showed renal detail previously unseen in QSM. CONCLUSION: Free-breathing QSM protocols in the human kidney using respiratory gating show equal susceptibility results compared to breath-hold acquisition while increasing clinical feasibility. A free-breathing protocol allows for the measurement of a larger subject pool and enables an increase in critical sequence parameters, such as echo-time and resolution. This could improve the accuracy of susceptibility calculations in the human abdomen.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An external reference to enable quantitative susceptibility mapping (QSM) in absolute terms: theoretical considerations, simulations, and experimental evaluation.Magma (New York, N.Y.) · 2026
- A geometric anthropomorphic phantom for quantitative susceptibility mapping: accuracy and repeatabilitybioRxiv · 2026
- Susceptibility source separation via Closed-form Alternating Newton-type OptimizationarXiv · 2026
- A physics-constrained Bayesian framework for QSM with uncertainty quantification and variational susceptibility source separation.Physics in medicine and biology · 2026
- Separation of Diamagnetic and Paramagnetic Susceptibility Sources in Liver Using DECOMPOSE‐QSMNMR in Biomedicine · 2026
- Respiratory rate estimation by using realignment data from a 1-minute high-temporal-resolution MRI sequence: A feasibility study.Acta radiologica (Stockholm, Sweden : 1987) · 2026