End-to-end differentiable volumetric modulated arc therapy optimization via computation graph-driven automatic differentiation.
Hongkai Wang, Wenzheng Sun, Wenlong Xia, Bin Liang
Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China. · Department of Radiation Oncology, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, Zhejiang, China.
Key points
From the paper's conclusion- The proposed differentiable VMAT framework demonstrated superior goal-directed optimization fidelity compared with MatRad-more reliably translating objectives into intended dosimetric outcomes-achieving plan quality approaching clinical Eclipse plans with faster optimization.
Abstract
BACKGROUND: Conventional volumetric modulated arc therapy (VMAT) optimization relies on gradient-based methods in which analytical gradients must be explicitly derived for each objective function and machine constraint. This dedicated mathematical derivation imposes substantial barriers to extending the optimization framework with new clinical objectives or delivery constraints, and the resulting CPU-based implementations do not natively exploit modern GPU hardware. PURPOSE: To develop an end-to-end differentiable VMAT optimization framework using computation graphs that enable automatic differentiation-based gradient computation for both fluence map optimization (FMO) and direct aperture optimization (DAO), while incorporating VMAT machine delivery constraints within a unified, GPU-accelerated pipeline. METHODS: A two-stage pipeline-fluence map optimization (FMO) followed by DAO-was developed. FMO reformulates dose calculation as a differentiable computation graph, enabling gradient computation for arbitrary objective functions including non-smooth Heaviside step terms. For DAO, a novel analytically derived computation graph maps MLC leaf positions to delivered fluence through closed-form time-averaged bixel exposure integration with finite-width boundary corrections, ensuring exact differentiability. Machine delivery constraints-including leaf collision, speed limits, and dose rate bounds-are incorporated as differentiable penalty terms for joint end-to-end optimization. Twenty lung cancer cases from the GDP-HMM AAPM Challenge dataset (60 Gy/30 fractions) were optimized using objectives derived from clinical Eclipse reference plans and compared against MatRad and Eclipse. RESULTS: The proposed method achieved PTV D98 of 58.88 ± 0.33 Gy, comparable to Eclipse (59.07 ± 0.55 Gy, p = 0.26) and significantly higher than MatRad (57.27 ± 0.48 Gy, p < 0.001). The Paddick conformity index was 0.81 ± 0.06, comparable to Eclipse (0.84 ± 0.22, p = 0.48) and superior to MatRad (0.64 ± 0.11, p < 0.001). OAR sparing closely matched Eclipse: heart mean dose 10.74 ± 4.89 Gy vs. 11.55 ± 4.76 Gy; lung mean dose 15.33 ± 3.50 Gy vs. 16.70 ± 3.29 Gy; LAD mean dose 9.55 ± 4.97 Gy vs. 9.05 ± 3.94 Gy (p = 0.18). MatRad achieved lower heart doses at the cost of degraded PTV coverage and conformity. Optimization time was significantly shorter (59.28 ± 43.95 s vs. 390.30 ± 184.90 s, p < 0.001), while delivery times were equivalent (p = 0.92). CONCLUSIONS: The proposed differentiable VMAT framework demonstrated superior goal-directed optimization fidelity compared with MatRad-more reliably translating objectives into intended dosimetric outcomes-achieving plan quality approaching clinical Eclipse plans with faster optimization.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- From Predicted Dose to Computationally Verified Plan: An End-to-End Deep Learning and Monte Carlo Pipeline for Automated VMAT Glioblastoma Radiotherapy.Journal of imaging informatics in medicine · 2026
- A virtual physicist framework integrating dose prediction and large language models for automated treatment planning.Journal of applied clinical medical physics · 2026
- Multi-modal direct LINAC parameter prediction for pancreatic VMAT: An optimization-free approach.Medical physics · 2026
- Quantum-Inspired Hamiltonian Descent for Radiation Therapy Treatment PlanningarXiv · 2026
- Prescription-aware diffusion Transformer for esophageal VMAT dose prediction: a retrospective dual-center validation and target-level analysisBioMedical Engineering OnLine · 2026
- Learning the Dose Manifold: A Latent Diffusion Framework for Volumetric Dose Prediction in Automated Radiotherapy Treatment PlanningarXiv · 2026