Daytime Idle Energy Reduction in Radiotherapy Simulation Computed Tomography: A Prospective Workflow-Based Evaluation.
Kouta Hirotaki, Yoshiaki Honda, Masashi Ito
Department of Radiological Technology, National Cancer Center Hospital East, Kashiwa, Chiba, Japan. · Department of Radiology, Saiseikai Niigata Kenoh Kikan Hospital, Niigata, Japan.
Key points
From the paper's conclusion- Power-saving operation reduced daytime energy consumption without meaningful workflow disruption. Radiotherapy simulation CT has prolonged, procedure-dependent idle periods arising from radiotherapy-specific preparation workflows, highlighting a practical and previously undercharacterised opportunity for energy reduction.
Abstract
INTRODUCTION: Previous computed tomography (CT) energy-saving studies have primarily focused on after-hours shutdown or idle energy use in diagnostic CT. Radiotherapy simulation, however, includes preparation and immobilisation steps that may prolong daytime non-acquisition periods. This study quantified daytime energy use in radiotherapy simulation CT and evaluated the contribution of power-saving operation. METHODS: In this prospective single-centre study, electrical power consumption was directly measured using a three-phase power logger during five baseline and five power-saving weekdays (08:00-18:00). Power measurements were integrated with patient in-room intervals and simulation procedure categories. Staff perceptions of workflow impact were assessed using an anonymous questionnaire, and carbon implications were estimated from the measured energy reduction. RESULTS: Mean directly measured clinical-hour energy consumption decreased from 38.7 to 36.5 kWh/day with power-saving operation, a 5.8% reduction. Non-acquisition time during patient presence varied markedly by procedure, from 56.5% for respiratory-gated four-dimensional CT to 97.1% for head-and-neck simulation. Power-saving operation shifted daytime power use towards a lower-power state, corresponding to an estimated reduction of 0.84-1.01 kg-CO2 per clinical day across the emission-factor sensitivity range. Twenty of 21 staff (95.2%) reported no impact on patient throughput. CONCLUSION: Power-saving operation reduced daytime energy consumption without meaningful workflow disruption. Radiotherapy simulation CT has prolonged, procedure-dependent idle periods arising from radiotherapy-specific preparation workflows, highlighting a practical and previously undercharacterised opportunity for energy reduction.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantifying energy savings in radiology: a simple approach to make the radiology department more sustainableBMJ Leader · 2025
- Automated linac downtime tracking across 4 radiation oncology clinics within a single hospital system utilizing statistical process control.Journal of applied clinical medical physics · 2026
- Energy consumption of standard and contrast-enhanced mammography: a step towards sustainable breast imagingEuropean Radiology · 2026
- Energy performance of MRI systems: on-site validation and comparison with manufacturer declarationsEuropean Radiology Experimental · 2026
- Greening computed tomography: a position paper on environmentally responsible imaging practice from the computed tomography subspecialty section of the Italian society of radiology.La Radiologia medica · 2026
- Energy sustainability of breast ultrasound: comparison of automated and handheld ultrasound systems.Ultrasonography (Seoul, Korea) · 2026