Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions

Clinical Characteristics, Mechanisms, and Management of Gore Cardioform Device-Related Thrombosis: A Comprehensive Review.

Aristides Plaitis, Nikolaos Arnas, Alexandros Archontikis, Tryfon Tzinieris, Stavroula Pantzartzi, Sarantos Linardakis, Nikitas Katsillis, Nikolaos Papakonstantinou,

Tzaneio General Hospital of Piraeus, Piraeus, Greece.

Review

Key points

From the paper's conclusion
  • Given the absence of device-specific guidelines, we propose a practical, clinically oriented framework for the prevention, detection, and management of suspected DRT in patients undergoing PFO or ASD closure with Gore Cardioform devices.

Abstract

Device-related thrombosis (DRT) is a rare but potentially serious complication following transcatheter closure of patent foramen ovale (PFO) and atrial septal defect (ASD). Although the Gore Cardioform family of septal occluders is increasingly used in contemporary structural heart interventions, evidence regarding associated thrombotic complications remains limited and fragmented. In this narrative review, we systematically summarize the available literature on DRT associated with Gore Cardioform devices, including randomized clinical trials, observational studies, and published case reports. The overall incidence appears low; however, reported cases highlight a heterogeneous clinical presentation, ranging from incidental imaging findings to systemic embolic events. Thrombus formation most commonly occurs within the early post-implantation period, although late events have also been described. Potential mechanisms are multifactorial and include incomplete endothelialization, flow disturbances related to device design, and patient-specific prothrombotic conditions. Identified risk factors include hypercoagulable states, atrial arrhythmias, inadequate antithrombotic therapy or poor adherence, large defect size, and device-related factors affecting septal apposition. Current management strategies are largely empirical and typically involve systemic anticoagulation, with most cases resolving without the need for invasive intervention. In selected high-risk or refractory cases, thrombolysis or surgical removal may be required. Given the absence of device-specific guidelines, we propose a practical, clinically oriented framework for the prevention, detection, and management of suspected DRT in patients undergoing PFO or ASD closure with Gore Cardioform devices.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Clinical Characteristics, Mechanisms, and Management of Gore Cardioform Device-Related Thrombosis: A Comprehensive Review — Science Skim