Balloon Rupture During Balloon-Expandable TAVR: Insights Into the Role of Calcium Distribution and Lesion Preparation From Two Cases.
Vagisha Sharma, Shiavax J Rao, Guy Rogers, Sudhakar Sattur, Edo Kaluski
Department of Cardiology, Guthrie Robert Packer Hospital, Sayre, Pennsylvania.
Key points
From the paper's conclusion- We present 2 cases demonstrating how calcium distribution and timing of rupture relative to valve expansion determine outcome.
Abstract
Balloon rupture during balloon-expandable transcatheter aortic valve replacement is a rare but potentially catastrophic complication, often driven by an underlying anatomic substrate. We present 2 cases demonstrating how calcium distribution and timing of rupture relative to valve expansion determine outcome. Late rupture after near-complete expansion resulted in no hemodynamic consequence, whereas early rupture in the setting of dense intravalvular calcification caused incomplete expansion, severe aortic regurgitation, ventricular fibrillation, repeat balloon failure, and fatal stroke. These cases highlight the importance of calcium assessment, lesion preparation, and anatomy-tailored procedural planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Longitudinal Deployment Balloon Rupture With Uncomplicated Retrieval During TAVR.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Gutter Rotation Technique for Percutaneous Ipsilateral Retrieval of a Ruptured SAPIEN-3 Balloon.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Bailout cardiac surgery in patients undergoing transcatheter aortic valve replacement: a comprehensive analysis of post-marketing safety reportsmedRxiv · 2026
- Prosthesis-to-Prosthesis Balloon Interaction During TAVR Causing Aorto-Mitral Curtain Injury and Left Ventricular Pseudoaneurysm.JACC. Case reports · 2026
- Remodel the Failed Trifecta or Not: That Is a Question.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Impact of basal septal bulge on conduction disturbance relative to valve position after transcatheter aortic valve replacementEuropean Heart Journal · 2026