Simultaneous Aortic and Mitral Valve-in-Valve Replacement Using Balloon-Expandable MyVal Valves With Bioprosthetic Valve Fracture.
Nitin Bajaj, Navreet Singh, Ankush Gupta, Sunny Pathania
Department of Cardiology, Army Institute of Cardiothoracic Sciences, Pune, India.
Key points
From the paper's conclusion- He underwent successful TDVR with a 20 mm MyVal balloon-expandable in the aortic position, followed by a post-implantation valve fracture, and a 27.5 mm MyVal transseptally in the mitral position.
Abstract
Valve-in-valve transcatheter aortic valve replacement and mitral valve replacement have proved to be a viable alternative to redo surgery in patients who have bioprosthetic valve dysfunction. There is a large subgroup of patients who undergo double valve replacement with bioprosthetic valves, especially in patients with rheumatic heart disease. Failure of these bioprosthetic valves would require a transcatheter management strategy. Transcatheter double valve replacement (TDVR) in these complex cases would require careful evaluation and planning of valve sizing, aortomitral interactions, access, and residual gradients. A 73-year-old man with rheumatic heart disease and prior double bioprosthetic valve replacement in 2014 presented with New York Heart Association class (NYHA) III dyspnea and exertional pre-syncope. He had severe stenosis of a 19 mm Perimount Magna aortic bioprosthesis and a 27 mm Perimount Magna mitral bioprosthesis, with pulmonary hypertension. Computed tomography demonstrated suitable anatomy for transcatheter valve-in-valve treatment. He underwent successful TDVR with a 20 mm MyVal balloon-expandable in the aortic position, followed by a post-implantation valve fracture, and a 27.5 mm MyVal transseptally in the mitral position. The final mean gradients were 15 mmHg across the aortic valve and 6 mmHg across the mitral valve at discharge. At 1-year follow-up, the patient remained clinically well without an increase in valve gradients. Simultaneous transcatheter aortic and mitral valve-in-valve replacement can be performed successfully in selected high-risk patients after meticulous multimodality planning. In patients with a small surgical aortic bioprosthesis, intentional bioprosthetic valve fracture may be an important adjunct to optimize post-implantation hemodynamics. This report highlights the complexity of procedural planning for bioprosthetic valve fracture in a small aortic valve, ensuring optimal hemodynamics and the safety and durability of the Myval transcatheter heart valve in challenging and complex procedures like TDVR.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Treatment of Failed Mitral Valve Bioprostheses or Repairs: Transcatheter Mitral Valve-in-Valve/Valve-in-Ring Versus Redo Surgical Mitral Valve Replacement.Journal of clinical medicine · 2026
- Valve-in-valve transcatheter mitral valve replacement via totally endoscopic endoclamp-assisted approach after failed bioprosthetic mitral valve: a case report.Journal of cardiothoracic surgery · 2026
- Staged Transcatheter Valve-in-Valve Replacement for Simultaneous Failure of 3 Surgical Bioprosthetic Valves.JACC. Case reports · 2026
- Single-Session Transcatheter Aortic and Tricuspid Valve-in-Valve Implantation in a Patient with Prior Multivalve Surgery and a Chronic Transvenous Pacing Lead: A Case-Based Review of Procedural Strategy.Life (Basel, Switzerland) · 2026
- Multivalvular heart disease and outcomes after transcatheter aortic valve implantation: updated results from the MIRA studyEuropean Heart Journal · 2026
- Valve‐in‐Valve Transcatheter Mitral Valve Replacement Versus Redo Surgical Mitral Valve Replacement: Meta‐Analysis of Early and Late OutcomesJournal of the American Heart Association · 2026