Staged Transcatheter Valve-in-Valve Replacement for Simultaneous Failure of 3 Surgical Bioprosthetic Valves.
Dominador V Del Castillo, Yan Wang
Division of Structural Heart, Xiamen Cardiovascular Hospital Xiamen University, Xiamen, China. Electronic address: dominador_delcastillo@yahoo.com. · Division of Structural Heart, Xiamen Cardiovascular Hospital Xiamen University, Xiamen, China.
Key points
From the paper's conclusion- Valve-in-valve transcatheter aortic valve replacement with a 23-mm supra-annular self-expanding valve and bioprosthetic valve fracture was performed first, followed by valve-in-valve (ViV)-transcatheter mitral valve replacement and ViV transcatheter tricuspid valve replacement with 26-mm balloon-expandable valves.
Abstract
OBJECTIVES: To describe a heart team-driven staged transcatheter strategy for accelerated triple bioprosthetic valve failure in a patient with prohibitive surgical risk. KEY STEPS: A 78-year-old woman with systemic lupus erythematosus, stage 4 chronic kidney disease, permanent atrial fibrillation, and severe frailty presented with NYHA functional class III heart failure. She had undergone triple surgical bioprosthetic heart valve replacement 8 years earlier. Imaging demonstrated severe aortic, mitral, and tricuspid bioprosthetic valve failure. Given prohibitive surgical risk, a staged transcatheter strategy was selected. Valve-in-valve transcatheter aortic valve replacement with a 23-mm supra-annular self-expanding valve and bioprosthetic valve fracture was performed first, followed by valve-in-valve (ViV)-transcatheter mitral valve replacement and ViV transcatheter tricuspid valve replacement with 26-mm balloon-expandable valves. All procedures were successful without complications. At 2 months, NYHA functional class improved from III to II, with satisfactory function of all 3 valves. POTENTIAL PITFALLS: Coronary obstruction, patient-prosthesis mismatch, neo-left ventricular outflow tract obstruction, and hemodynamic intolerance. TAKE-HOME MESSAGES: Staged triple ViV intervention may provide a feasible alternative to redo surgery. Multimodality imaging and individualized sequencing are critical.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Simultaneous Aortic and Mitral Valve-in-Valve Replacement Using Balloon-Expandable MyVal Valves With Bioprosthetic Valve Fracture.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- TAV-in-TAV-in-SAV in a Patient With Prior Mitral Valve-in-Valve.JACC. Case reports · 2026
- Single-Session Transcatheter Aortic and Tricuspid Valve-in-Valve Implantation in a Patient with Prior Multivalve Surgery and a Chronic Transvenous Pacing Lead: A Case-Based Review of Procedural Strategy.Life (Basel, Switzerland) · 2026
- Percutaneous Tricuspid Valve-in-Valve Implantation With Melon Seeding During Deployment.JACC. Case reports · 2026
- Severe Bioprosthetic Mitral Valve Dysfunction: Multimodality Imaging Identifies Structural Valve Degeneration and Guides Valve-in-Valve TMVR.JACC. Case reports · 2026
- Treatment of Failed Mitral Valve Bioprostheses or Repairs: Transcatheter Mitral Valve-in-Valve/Valve-in-Ring Versus Redo Surgical Mitral Valve Replacement.Journal of clinical medicine · 2026