Biomedical Chromatography

Development and Validation of a Stability‐Indicating HPLC Method for Degarelix and Characterization of Its Degradation Impurities Using LC–MS, 1 H NMR, and FTIR Techniques

Pydi Simhachalam, Thatituri Sabithakala

Key points

From the paper's conclusion
  • A new stability‐indicating RP‐HPLC method was developed and validated for Degarelix (DGL) and its impurities, and the degradation characteristics of the drug under various stress conditions were systematically evaluated.

Abstract

A new stability‐indicating RP‐HPLC method was developed and validated for Degarelix (DGL) and its impurities, and the degradation characteristics of the drug under various stress conditions were systematically evaluated. Forced degradation experiments showed that DGL degraded significantly under acidic (5.281%), alkaline (2.0%), and thermal (2.950%) conditions and, to a lesser extent, under oxidative (1.634%), photolytic (1.377%), and humidity (1.400%) conditions. This separation process was performed on an AURASHELL C18 PFP (150 × 4.6 mm, 2.7 μm) column using a gradient elution method with sodium perchlorate buffer (pH 4.4), acetonitrile, and isopropyl alcohol; detection was carried out at a wavelength of 225 nm. The developed method successfully separated DGL, achieving good peak symmetry and resolution between DGL and its impurities. LC–MS, 1H NMR, and IR spectroscopy were used to characterize the degradation products (descarbamoyl DGL and N‐carbamoyl DGL) and to confirm their respective structural changes. In accordance with ICH sensitivity guidelines, the method was validated with LOD values of 0.031–0.034 μg/mL and LOQ values of 0.1147–0.2480 μg/mL. This method is simple and exhibits correlation coefficients ( R 2 ) greater than 0.996. This method also demonstrated the robustness and stability of that solution under test conditions.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Development and Validation of a Stability‐Indicating HPLC Method for Degarelix and Characterization of Its Degradation Impurities Using LC–MS, 1 H NMR, and FTIR Techniques — Science Skim